OP_medinCNO EN
Rev 12 Stand 06.03.2019
25
von
68
Figure 4-2: Battery charge status displayed in the top left-hand area of the screen
The charge status of the internal battery is displayed in the top left-hand area of the medinCNO
®
screen.
Attention:
When the medinCNO
®
is running on battery power, an alarm is sounded when the internal battery is low. It
must then be immediately connected to an external power source, as otherwise the medinCNO
®
will stop
working.
To disconnect the medinCNO
®
from the mains, the plug of the power supply unit of the medinCNO
®
must be
pulled out of the electrical outlet. Otherwise the medinCNO
®
is still connected to the mains, even when it is
turned off.
Therefore, when setting up the medinCNO
®
, it should be ensured that the plug of the power supply unit is still
accessible even when set up and the medinCNO
®
can thus be easily disconnected from the mains.
Disconnect the power supply unit from the mains during thunderstorms.
Warning:
Since the surface temperature of the power supply unit can become too hot for the patient, it should always
be ensured that the power supply unit is not within the patient's reach.
The plug of the power supply unit of the medinCNO
®
must also always be freely accessible even when set up,
in order to allow the medinCNO
®
to be quickly disconnected from the mains in the event of a hazardous
situation.
Do not insert the power supply plug of the medinCNO
®
into ceiling outlets since, due to its weight, it can be
pulled out of the outlet in this case.