5.2
Display screen
Figure 5-2: Display and button assignment of the medinCNO
®
a) Flow; b) FiO
2
value or start and
discontinuation of the
O
2
flush; c) Additional push flow; d) Manual push trigger button; e) Readings; f) Mode;
g) Status displays;
h) Pressure course graph; i) Alarm messages; k) Alarm reset button; m) to r) Buttons for other settings
The medinCNO
®
display screen shows the values set, the readings, status messages and the names of the
buttons surrounding the screen:
-
Mode
: displays the selected mode - position f in Figure 5-2
-
Basic CPAP settings:
the target values set for
flow
,
FiO
2
and the additional push flow selected are
shown on the right of the display screen - Figure 5-2 position a (
Flow
), b (
FiO
2
) and c (additional
push flow -
F
insp
) After pressing button b twice, the setting of the
O
2
flush
and its start button are
indicated instead of the set target value for
FiO
2
.
-
Manual push:
the manual push trigger button, position d in Figure 5-2.
-
Readings
: the values measured by the medinCNO
®
for the average CPAP pressure (
CPAP
) and the
measured oxygen concentration (
FiO
2
). In addition, mode-dependent readings such as the
respiratory (
RR
) or push frequencies (
R
insp
- corresponds to the number of inspiration pushes per
minute automatically emitted by the medinCNO
®
) are shown above in the coloured bar of the
medinCNO
®
(Figure 5-2 position e)
-
Status displays:
the medinCNO
®
displays battery status and interrupted alarm sounds at position g
-
Pressure course curve
: the CPAP pressure measured over time is shown as a graph in position h in
Figure 5-2 . In addition, the maximum measured pressure for each push is marked numerically. The
representation of the pressure course can be helpful in selecting CPAP parameter settings.
The pressure curve can be indicated on a scale of 0-10, 0-15 or 0-20 mbar/cmH
2
O.
-
Active alarms:
information about active alarms is given in position i in Figure 5-2. If several alarms
are active at the same time, they are displayed in turn. See chapter 8 for a detailed description of
alarms and tips on how to avoid them occurring.