Check gas feed pressure air
red/
high
3s
The air feed pressure is too low (<3 bar) or
too high (>6.0 bar) - the medinCNO
®
is only
providing oxygen.
This alarm starts the emergency
programme.
CPAP pressure high
yellow/
medium
<87s
(3)
The CPAP average pressure was
continuously greater for >1 second than
the set upper CPAP pressure alarm
threshold.
- In a normal case, the average value
interprets the CPAP readings from the past
five seconds which delays the start of the
alarm up to <8 seconds.
- If the backup is activated, there may be a
maximum delay of 85 seconds until the
average value has been fully updated,
which can result in a total alarm delay of
<87 seconds.
- In SNIPPV mode without active backup,
the average value is updated every 15
seconds, and for this reason, there is a
total alarm delay of <17 seconds.
In this case, the medinCNO
®
continues to
work normally.
CPAP pressure low
yellow/
medium
<101s
(2)(3)
For 15 seconds, the CPAP average pressure
was constantly less than the set lower
CPAP pressure alarm threshold.
- In a normal case, the average value
interprets the CPAP readings from the past
five seconds which delays the start of the
alarm up to <22 seconds.
- If the backup is activated, there may be a
maximum delay of 85 seconds until the
average value has been fully updated,
which can result in a total alarm delay of
<101 seconds.
- In SNIPPV mode without active backup,
the average value is updated every 15
seconds, and for this reason, there is a
total alarm delay of <31 seconds.
In this case, the medinCNO
®
continues to
work normally.
Oxygen concentration high
yellow/
medium
100s
The oxygen concentration measured is
higher than the alarm threshold set for
~1.5 min - the medinCNO
®
continues to
operate normally.
Oxygen concentration low
yellow/
medium
100s
The oxygen concentration measured is
higher than the alarm threshold set for
~1.5 min - the medinCNO
®
continues to
operate normally.