3.3
Gas feed
In order to operate the medinCNO
®
, it must be supplied with air and oxygen from an external source that meets
these conditions:
Medicinal oxygen
:
-
300 to 600 kPa (=3.0 to 6.0 bar)
-
Oxygen content: ≥ 99.5%
-
H
2
O ≤ 67 ppm (V/V) (
free from condensate)
Medicinal air (Aer medicalis):
-
300 to 600 kPa (=3.0 to 6.0 bar)
-
Oil ≤ 0.1 mg/m
3
, ideally oil-free
-
H
2
O ≤ 870 ppm (V/V) and free from condensate
! If as a result of the operating temperature it is
possible that condensate may form despite a relatively low concentration of H
2
O in the air, then
the concentration of H
2
O must be reduced accordingly.
Figure 3-3: Water trap in the medinCNO
®
air inlet; a) Bolt for removing the condensate from the water trap
The medinCNO
®
can be fitted with a water trap in the air inlet as an option (Figure 3-3) (REF 39-800). This allows
small amounts of condensate to be removed from the air supply provided to the medinCNO
®
if the water trap
is checked for condensate and emptied on a regular basis. To empty the water trap, undo the black bolt (Figure
3-3 – a) on the underside of the water trap and keep it open until the condensate has run off (this can also be
done under pressure).