Temperature
yellow/
medium
90s
The temperature in the medinCNO
®
is too
high. Check the environmental conditions
and/or the prefilter and contact a service
technician.
Connect power supply unit -
battery low
yellow/
medium
6s
The internal battery is almost discharged.
The power supply unit must be connected
to the mains immediately.
Fan blocked
yellow/
low
15s
The fan is blocked, please contact a service
technician.
Leak-Assist
yellow/
low
<36s
This alarm is triggered if the average CPAP
is different from the target CPAP for 30
seconds and if the leak flow of ±2 L/min is
not sufficient to compensate.
In this case, the medinCNO
®
continues to
work normally.
Pure alarm sound, medinCNO
®
switched off
-
<1s
This alarm is triggered if the medinCNO
®
power supply fails completely, for example
because the power supply unit was not
connected to the mains although the
battery was discharged.
The medinCNO
®
was switched off.
Press Alarm Reset to restart flow
red/
high
-
No alarm – informational message only
Triggered by a prolonged high feed
pressure (e.g. if a flow tube is blocked), by
several high CPAP pressures successively or
by a combination of high feed pressures
and high CPAP pressures, the flow supply is
interrupted for the time being and is only
reactivated after the Alarm Reset button is
pressed. Reactivation of the flow takes
approximately 10 seconds starting from
this moment.
Push activated
turquoise/
info
-
Not actually an alarm, just notification that
a push was triggered. Users can decide
whether or not to receive this information
visually and aurally.
Apnoea – push activated
turquoise
/
info
-
Not actually an alarm, just a notification
that a push was triggered due to apnoea.
Users can decide whether or not to receive
this information visually and aurally.
Backup Push
gray/
info
-
Not actually an alarm, just a notification
that a backup push was triggered due to
apnoea. Users can decide whether or not
to receive this information aurally.
However, this information is always
indicated visually.
(1) Delay ∆t: This corresponds to the
maximum time which passes from the time the alarm condition occurs until the alarm is
triggered.