+49 8142 448 460
www.medin-medical.com
medin Medical Innovations GmbH
Adam-Geisler-Str. 1
D-82140 Olching
OP_medinCNO EN
Rev 12 Stand 06.03.2019
10
von
68
3
Installation of the medinCNO
®
– Conditions
Prior to start-up, the medinCNO
®
must be securely mounted. In doing so, the following conditions must be
observed:
3.1
Mounting
Figure 3-1: a) Brackets holding the medinCNO
®
in position; b) Ventilation slits; c) Overpressure valve; d)
Knurled screws
The medinCNO
®
must always be securely mounted before the device is started up. To do this, the two fixing
brackets on the back of the device (Figure 3-1-a) are inserted into a standard rail (cross-section: 10 x 25 mm)
and the brackets are then fixed in place manually using the knurled screws (Figure 3-1– d). It is important to
ensure that the rear and lateral ventilation slits (Figure 3-1 – b), including prefilters, are not covered and that
the overpressure valve (Figure 3-1– c) on the base of the device is not blocked.
The medinCNO
®
can be transported when it is not in use (no connection to the patient). When it is in use, it
must be securely mounted. However, if the medinCNO
®
becomes damaged during transport or due to rough
handling, its function must be checked by a service technician. It is not allowed to be used until then.
Warning:
-
The medinCNO
®
must always be securely mounted in an upright position throughout the entire
time that it is in use.
-
The medinCNO
®
must not be covered while in use, and none of the openings or ventilation slits
must be blocked.
-
The overpressure valve must be kept clear whenever the medinCNO
®
is in use, and it must always
hang free in the air.
-
A damaged medinCNO
®
may not be used.