Attention:
The functionality of the medinCNO
®
can only be guaranteed if both the oxygen and medicinal air supplied to it
are free from condensate water, oil and contaminants. If this is not guaranteed, the medinCNO
®
could be
damaged. Even the use of a water trap can only remove small amounts of condensate from the air.
In order to avoid fires, oxygen must not come into contact with oil or grease. It is therefore essential to ensure
that all parts of the medinCNO
®
and the CPAP system which come into contact with oxygen (e.g. gas feed tubes,
O
2
inlet to the medinCNO
®
, patient outlet, patient tubes and Medijet
®
) remain free from oil and grease.
Gas feed connections:
The medinCNO
®
can be fitted as standard with NIST or DISS gas supply connections as the customer prefers.
The tube kits to connect the device to the external gas source must also meet the selected standard.
3.4
External alarm interface and data interfaces
Attention:
Only devices and systems which are approved for medical use and which meet the relevant electrical safety
and EMC standards may be connected to the medinCNO
®
data interface and alarm interface.
3.4.1
External alarm
The medinCNO
®
has an interface for connection to an external alarm system. The interface can open or close
an external current circuit in the event of an alarm. The interface thus acts only as a switch, that is, the power
supply must be ensured by the external alarm system.
Time delay:
The interface to the external alarm system is updated every 100 msec. That is, the alarm in the external alarm
system is activated with a maximum delay of 100 msec in comparison to the alarm in the device.
The alarm is likewise deactivated with a maximum delay of 100 msec in comparison to the alarm in the device.
The silencing of the alarm sound does not count here, but rather the actual deactivation of the alarm (that is,
correction of the alarm condition and acknowledgement of the alarm by the user).
The cable supplied by Medin must be used to connect the medinCNO
®
to an external alarm system in order to
achieve EMC according to DIN EN 60601-1-2:2016 (REF 39-114) and the connection must be set up initially by
an appropriately trained technician.
Attention:
The medinCNO
®
may only be connected to an external alarm system using the connection cable provided by
the manufacturer (REF 39-114), otherwise the functionality and EMC of the device cannot be guaranteed.
Warning:
The function of the external alarm interface must be initially activated by a service technician in the service
menu prior to its first use. The medinCNO
®
s are delivered with a deactivated interface as a standard feature.