(2) After start-up, the disconnection and CPAP pressure low alarm is suppressed for the first 5 minutes of operation. However,
the alarms are activated no later than after 5 min or if a CPAP pressure >1 mbar for at least 30 sec has already been measured.
(3) Maximum time delay, including maximum required time to completely re-calculate the average value.
The alarm volume is:
Type of alarm
Volume in (dB)
In the case of A-weighted
background noise volume (dB)
High priority (red)
69
53
Medium priority (blinking yellow)
67
52
Low priority (steady yellow)
67
52
Information signal (push)
68
53
Alterable alarms:
The user can set the
CPAP pressure low, CPAP pressure high
(yellow) alarms and the permitted deviation
interval for the
oxygen concentration high / low
alarms. This can be done using the setting in the
Alarm Settings
menu (see chapter 5.12). After the numerical values are adjusted, the modified settings become active and are
saved when this change is confirmed by pressing the dial. No other alarms can be modified.
Alarm settings upon start-up:
Upon start-up, the
CPAP pressure high
(yellow) alarm is set to 12 mbar and the
CPAP pressure low
alarm is set
to 3 mbar. The alarm interval for monitoring the oxygen levels set is always set to ±5% at start-up.
If the medinCNO
®
is restarted, the alarms are reset to these levels. This cannot be changed by the user.
The alarms -
CPAP pressure low
(yellow) and
CPAP pressure high
(yellow) analyse whether the average value of
the CPAP readings is outside of the alarm thresholds (can be set under Alarm Settings – see chapter 5.12). Thus
the alarm
CPAP pressure high
is triggered if the CPAP average value was above the alarm threshold for more
than one second. The alarm
CPAP pressure low
is triggered if the CPAP average value was below the alarm
threshold for more than 15 seconds.
In oscillation mode, the
Disconnection
alarm is triggered if the average CPAP value was less than 2 mbar for
five seconds.
Since the
Disconnection
(only in oscillation mode),
CPAP pressure low
and
CPAP pressure high
alarms refer to
the average CPAP pressure, the start and end of the alarm is thus further delayed (see table at the start of the
chapter). This likewise causes the
CPAP pressure low
or
Disconnection
alarm during oscillation to not be reset
until the CPAP average value is again above the alarm threshold and not directly after adjusting the alarm
threshold or correcting the problem.
The alarm
oxygen concentration high/low
is triggered when the measured oxygen concentration is outside of
the set limits. These thresholds are always automatically set by the medinCNO
®
in
automatic FiO
2
alarm
thresholds
mode at the level of the set interval (can be set under alarm settings; see chapter 5.12) around the
target oxygen concentration. In
FiO
2
alarm thresholds manual
mode, they are set via manual updating at the
level of the set intervals around the oxygen target value active at this time. The thresholds that are active due
to the current setting are displayed in the target oxygen concentration field as superscript and subscript
numerals. Thus, for example, at a set target oxygen concentration of 30% and an alarm interval setting of ±3%,
the
oxygen concentration low
alarm is triggered if the measured oxygen concentration is less than 27%, and
the
oxygen concentration high
alarm is triggered if the measured oxygen concentration is greater than 33%.