Please note:
If, due to an alarm, there is a high CPAP pressure (red), feed pressure high (or Press Alarm Reset to restart flow)
or a disconnection alarm for an interruption in the oscillation, the values originally set for amplitude and flow
must not be reset but instead the values will be gradually reverted to.
In the case of all other higher-priority alarms (red), the values for amplitude and flow must be reset!
5.11
Switching between modes
Pressing button m in Figure 5-2 opens mode selection. In this case the available modes are shown at the bottom
of the screen. The user can now select the desired mode by pressing the corresponding button. It is important
to note that CPAP mode and Apnoea CPAP mode are operated by the same button. Pressing this button
repeatedly allows the user to switch between the two modes. The selected mode appears on a blue
background. The mode selected must then be confirmed by pressing the
Confirm
button. Only then does the
medinCNO
®
switch to this mode.
The user can check whether mode switch has taken place correctly by checking the mode display in the top
left-hand area of the screen after making the switch.
If the NIPPV special case is active in Apnoea CPAP mode, this is displayed instead of the mode display.
When switching between two modes, the medinCNO
®
transfers all the settings that are available in both
modes.
The following situations represent exceptions:
-
Oscillation mode: If this ends, the set flow value is only transferred if it is ≤
8 L/min. Otherwise it is
reduced to 8 L/min.
-
Apnoea CPAP: If sensitivity is deactivated, this is adjusted in the event of a change in the SNIPPV to
a value of ±0.5.
Please note:
If a highest-priority alarm (e.g. high pressure or lack of gas) and the safety mode triggered as a result are active,
the mode may not be changed. The mode can only be changed once again after the error has been corrected.
A mode change during the safety mode is only possible if the safety mode was activated by the
Disconnection
alarm
.
5.12
Alarm settings
The alarm sub-menu is available in each medinCNO
®
mode. This sub-menu can be opened by pressing button
The points described below can be changed after opening this sub-menu. If two functions are allocated to a
button, the operator can press the relevant button repeatedly to switch between the two functions. Individual
values are changed by selecting the corresponding value (visible thanks to its blue background), then using the
dial to change the value and finally confirming the new value by pressing the dial.
CPAP high:
this corresponds to the freely selectable upper CPAP pressure alarm. If this value is exceeded by
the CPAP average value for a certain amount of time, a
CPAP pressure high
alarm is triggered after a delay (see
chapter 8.2).