8.2
Error messages during operation
Error message text
Colour/
Priority
∆
t
(1)
Meaning and device response
Disconnection (CPAP, Apnoea
CPAP, SNIPPV mode)
red/
high
<6s
(2)
The CPAP pressure is constantly <2.0 mbar
for >5 seconds - the flow tube or pressure
meter tube are no longer connected to the
medinCNO
®
. Please reconnect the tube.
This alarm starts the emergency
programme.
Disconnection (oscillation mode)
red/
high
<12s
(2) (3)
The CPAP pressure is constantly <2.0 mbar
for >5 seconds - the flow tube or pressure
meter tube are no longer connected to the
medinCNO
®
. Please reconnect the tube.
The average value interprets the CPAP
readings from the past five seconds which
delays the start and end of the alarm up to
12 seconds.
This alarm starts the emergency
programme.
CPAP pressure high
red/
high
<1s
CPAP pressure >18 mbar - the medinCNO
®
briefly interrupts the flow.
This alarm starts the emergency
programme.
Feed pressure high
red/
high
<1s
Feed pressure >40 mbar, the flow supply
tube is blocked or is generating resistance
that is too high. Please check. - The
medinCNO
®
briefly interrupts the flow.
This alarm starts the emergency
programme.
Device error - patient pressure.
Device must be replaced!
red/
high
<1s
The patient pressure sensor is causing an
error. The two CPAP sensors are measuring
a difference of >2 mbar for 500
milliseconds. The device can no longer be
used and must be replaced. Please contact
a service technician.
This alarm starts the emergency
programme.
Device error - device shut down
necessary
red/
high
<1s
A serious error occurred within the
medinCNO
®
. The medinCNO
®
can no longer
be used and must be replaced. Contact a
service technician.
This alarm starts the emergency
programme.
Check gas feed pressure O
2
red/
high
3s
The O
2
feed pressure is too low (<3 bar) or
too high (>6.0 bar) - the medinCNO
®
is only
providing air.
This alarm starts the emergency
programme.