Alarms and Error Messages
8.1
Error messages during system start-up
Error message text
Meaning
Connect power supply unit - battery low
The internal battery is almost discharged. Please connect
the power supply unit to the mains.
Fan blocked
The fan is blocked, please contact a service technician.
Gas feed pressure low, check O
2
The oxygen feed pressure is too low (<3 bar). Please
check oxygen feed.
Gas feed pressure high, check O
2
The oxygen feed pressure is too high (>6 bar). Please
check oxygen feed.
Gas feed pressure low, check air
The air feed pressure is too low (<3 bar). Please check air
feed.
Gas feed pressure high, check air
The air feed pressure is too high (>6 bar). Please check
air feed.
Oxygen calibration failed, check O
2
The oxygen calibration could not be performed correctly.
Please check whether the oxygen cell is working properly
and whether the medinCNO
®
is correctly connected to
the oxygen feed. The medinCNO
®
cannot be operated
without oxygen.
Device error - device shut down necessary
There is a defect in the device’s pneumatic system, at
the overpressure valve or rechargeable battery. The
medinCNO
®
can no longer be used and must be shut
down. Please contact a service technician.
Error during tube test - restart
An error occurred during the tube test (e.g. leakage).
Please repeat the tube test.
Device error - patient pressure. Device must
be replaced
The patient pressure sensor is causing an error. The
device cannot be used and must be replaced. Please
contact a service technician.
3V battery empty - battery has to be
exchanged.
The internal 3V battery is empty and must be replaced.
Please contact a service technician.
Battery needs replacing - device function in
battery mode cannot be guaranteed. Device
must be connected to mains for use!
The internal rechargeable battery is too old or was used
too frequently and must be replaced. Do not use the
medinCNO
®
on rechargeable battery power and contact
a service technician.
Maintenance necessary!
Next maintenance necessary:
YY.MM.DD
The last maintenance of the medinCNO
®
took place at
least 11 months ago and must be repeated no later than
the date indicated. Contact a service technician.