WR6K-OM-E Rev A
ISSUED: October 2003
45
Note:
You can set the granularity (delta) of the coarse adjustment by double-tapping inside the data entry field, then
touching the
Advanced
checkbox in the pop-up numeric keypad. The keypad presents
Coarse
delta up/down buttons to
set the delta:
.
In the pop-up keypad, be sure to leave the
Fine
checkbox unchecked to adjust the coarse delta.
Trigger Knobs:
Level
Selects the trigger threshold level. Press this knob to quickly set the level to
zero. The Level is indicated in the
Trigger
descriptor label
.
Trigger
Buttons:
Stop
Prevents the scope from triggering on a signal. If you boot up the instrument
with the trigger in Stop mode, the message "no trace available" will be
displayed. Press Auto to display your trace.
Auto
Triggers the scope after a time-out, even if the trigger conditions are not met.
Normal
Triggers the scope each time a signal is present that meets the conditions set
for the type of trigger selected.
Single
Arms the scope to trigger once (single-shot acquisition) when the input signal
meets the trigger conditions set for the type of trigger selected. If the scope is
already armed, it will force a trigger.
Horizontal
Knobs:
Delay
Horizontally positions the scope trace on the display so you can observe the
signal prior to the trigger time.
Delay
adjusts the pre- and post-trigger time.
Press this knob to quickly set the delay to zero. The trigger point is positioned in
the middle of the display grid.
Time/Division
Sets the time/division of the scope timebase (acquisition system).
Содержание 6000 SERIES
Страница 59: ...WR6K OM E Rev A ISSUED October 2003 57 4 Select a language 5 Click Next when the Wizard appears ...
Страница 60: ...Operator s Manual 58 ISSUED October 2003 WR6K OM E Rev A 6 Select Complete setup 7 Click Install ...
Страница 65: ...WR6K OM E Rev A ISSUED October 2003 63 Proper Orientation of Drive ...
Страница 131: ...WR6K OM E Rev A ISSUED October 2003 129 Example ...
Страница 141: ...WR6K OM E Rev A ISSUED October 2003 139 ...
Страница 236: ...Operator s Manual 234 ISSUED October 2003 WR6K OM E Rev A Example 6 ...
Страница 247: ...WR6K OM E Rev A ISSUED October 2003 245 In this figure the panel has been pasted onto the Excel sheet for comparison ...
Страница 261: ...WR6K OM E Rev A ISSUED October 2003 259 ...
Страница 286: ...Operator s Manual 284 ISSUED October 2003 WR6K OM E Rev A ...
Страница 304: ...Operator s Manual 302 ISSUED October 2003 WR6K OM E Rev A ...
Страница 307: ...WR6K OM E Rev A ISSUED October 2003 305 Convolve two signals ...
Страница 327: ...WR6K OM E Rev A ISSUED October 2003 325 How JitterTrack s Interval Error works when Clock Mode is selected ...
Страница 348: ...Operator s Manual 346 ISSUED October 2003 WR6K OM E Rev A ...
Страница 398: ...Operator s Manual 396 ISSUED October 2003 WR6K OM E Rev A Setting Up the Line Voltage and Current Signals ...
Страница 402: ...Operator s Manual 400 ISSUED October 2003 WR6K OM E Rev A ...