To reduce noise:
The example at left shows enhanced
resolution of a noisy signal. The original trace (left top)
has been processed by a 2-bit enhanced resolution filter.
The result (left bottom) shows a "smooth" trace, where
most of the noise has been eliminated.
Note:
Enhanced resolution can only improve the resolution of a trace; it cannot improve the
accuracy or linearity of the original quantization. The pass-band will cause signal attenuation for
signals near the cut-off frequency. The highest frequencies passed may be slightly attenuated.
Perform the filtering on finite record lengths. Data will be lost at the start and end of the
waveform: the trace will be slightly shorter after filtering. The number of samples lost is exactly
equal to the length of the impulse response of the filter used: between 2 and 117 samples.
Normally this loss (just 0.2 % of a 50,000 point trace) is not noticed. However, you might filter a
record so short there would be no data output. In that case, however, the instrument would not
allow you to use the ERES feature.
To Set Up Enhanced Resolution (ERES)
1.
In the menu bar, touch
Math
, then
Math Setup...
in the drop-down menu.
2.
Touch a function tab
F1
through
Fx
The number of math traces available depends on the
software options loaded on your scope. See Specifications..
3.
Touch inside the
Operator1
data entry field.
4.
Select
ERES
from the
All Functions
or
Filter
group of Math functions.
5.
Touch the
Trace On
checkbox.
6.
Touch the "ERES" tab in the right-hand dialog, then touch inside the
bits
field and make
an "Enhance by" selection from the pop-up menu:
Содержание 6000 SERIES
Страница 59: ...WR6K OM E Rev A ISSUED October 2003 57 4 Select a language 5 Click Next when the Wizard appears ...
Страница 60: ...Operator s Manual 58 ISSUED October 2003 WR6K OM E Rev A 6 Select Complete setup 7 Click Install ...
Страница 65: ...WR6K OM E Rev A ISSUED October 2003 63 Proper Orientation of Drive ...
Страница 131: ...WR6K OM E Rev A ISSUED October 2003 129 Example ...
Страница 141: ...WR6K OM E Rev A ISSUED October 2003 139 ...
Страница 236: ...Operator s Manual 234 ISSUED October 2003 WR6K OM E Rev A Example 6 ...
Страница 247: ...WR6K OM E Rev A ISSUED October 2003 245 In this figure the panel has been pasted onto the Excel sheet for comparison ...
Страница 261: ...WR6K OM E Rev A ISSUED October 2003 259 ...
Страница 286: ...Operator s Manual 284 ISSUED October 2003 WR6K OM E Rev A ...
Страница 304: ...Operator s Manual 302 ISSUED October 2003 WR6K OM E Rev A ...
Страница 307: ...WR6K OM E Rev A ISSUED October 2003 305 Convolve two signals ...
Страница 327: ...WR6K OM E Rev A ISSUED October 2003 325 How JitterTrack s Interval Error works when Clock Mode is selected ...
Страница 348: ...Operator s Manual 346 ISSUED October 2003 WR6K OM E Rev A ...
Страница 398: ...Operator s Manual 396 ISSUED October 2003 WR6K OM E Rev A Setting Up the Line Voltage and Current Signals ...
Страница 402: ...Operator s Manual 400 ISSUED October 2003 WR6K OM E Rev A ...