WR6K-OM-E Rev A
ISSUED: October 2003
369
procedure contained in the Matching the Time Delay in Your Measurement System to
correct any delay between the current and voltage.
2.
Connect the voltage and current probes to the appropriate points in the circuit under test.
The figure above can be used as a guide. If measurements are to be made on the device
as a function of an event such as load change or start-up, select a signal to be used as a
trigger for this event. The example in the figure uses the DSO’s Channel 4 to acquire a
trigger signal indicating when the load changes from maximum to minimum. The DSO’s
EXT trigger input also could be used.
Preliminary Trigger Setup
Before you enter the PowerMeasure software menu, it is important to determine the source and
setup of the triggers. Identify the signal on which the main measurement will be triggered, as well
as the signal on which the acquisition of an extended measurement record is to be triggered. The
main trigger can be the device’s voltage or current signal while the event trigger is usually
associated with load change or turn-on and turn-off. Establish the event trigger first (if required),
then set up the main trigger. Choose the desired trigger signal and establish a stable display.
Typical Connections to Circuit under Test for MAIN and EVENT Triggers
Initial Setup (Optional Event Trigger)
Determine the event whose change in operation requires the recording of an extended data
record. Examples of such an event include the change from maximum to minimum load, and
start-up. Presetting the trigger of such an event will make the final power measurement setup
easier.
Содержание 6000 SERIES
Страница 59: ...WR6K OM E Rev A ISSUED October 2003 57 4 Select a language 5 Click Next when the Wizard appears ...
Страница 60: ...Operator s Manual 58 ISSUED October 2003 WR6K OM E Rev A 6 Select Complete setup 7 Click Install ...
Страница 65: ...WR6K OM E Rev A ISSUED October 2003 63 Proper Orientation of Drive ...
Страница 131: ...WR6K OM E Rev A ISSUED October 2003 129 Example ...
Страница 141: ...WR6K OM E Rev A ISSUED October 2003 139 ...
Страница 236: ...Operator s Manual 234 ISSUED October 2003 WR6K OM E Rev A Example 6 ...
Страница 247: ...WR6K OM E Rev A ISSUED October 2003 245 In this figure the panel has been pasted onto the Excel sheet for comparison ...
Страница 261: ...WR6K OM E Rev A ISSUED October 2003 259 ...
Страница 286: ...Operator s Manual 284 ISSUED October 2003 WR6K OM E Rev A ...
Страница 304: ...Operator s Manual 302 ISSUED October 2003 WR6K OM E Rev A ...
Страница 307: ...WR6K OM E Rev A ISSUED October 2003 305 Convolve two signals ...
Страница 327: ...WR6K OM E Rev A ISSUED October 2003 325 How JitterTrack s Interval Error works when Clock Mode is selected ...
Страница 348: ...Operator s Manual 346 ISSUED October 2003 WR6K OM E Rev A ...
Страница 398: ...Operator s Manual 396 ISSUED October 2003 WR6K OM E Rev A Setting Up the Line Voltage and Current Signals ...
Страница 402: ...Operator s Manual 400 ISSUED October 2003 WR6K OM E Rev A ...