Operator's Manual
366
ISSUED: October 2003
WR6K-OM-E Rev A
The signal time skew will be shown as delay difference between the voltage and current
waveform. By changing the value of deskew (either by directly entering a number using the pop-
up keypad or by adjusting the number with the
Adjust
knob), the current waveform will be moved
to the right (positive deskew) or left (negative deskew) to a maximum of +/- 100 ms. In this
example the delay difference (time skew) is a little more than 10 ns per division. The amount of
delay will depend on which voltage and current probes are used, as well as on the length of the
probes and the coaxial cable used to connect the DA1855A to the DSO.
Note:
When a DA1855A Differential Amplifier (connected by a 1.2 m coaxial cable), DXC100A
Differential Passive Probe Pair, and AP015 Current Probe are used, the time delay is matched
and no deskew adjustment is needed.
1.
Adjust the deskew value until the current waveform coincides with the voltage waveform.
If the current waveform’s fall time is slower than that of the voltage, align the beginning of
the waveforms (alignment point).
2.
Turn the
Adjust
knob until the knee of the current and voltage waveforms are aligned.
Use the
Fine
button under the
Adjust
knob to make small adjustments.
3.
Touch inside the
Deskew
field and enter a value of zero on the pop-up keypad to return
the Deskew value to zero. The time delays of these two channels are now matched, and
they can be used to make accurate measurements that require precise time alignment of
the current and voltage waveforms. This deskew value is only valid for this particular
setup. Changing probes or bandwidth for either the current or voltage channel requires
the channel’s time-delay difference to be checked and possibly corrected. This process
can be used to characterize and correct the delay difference between more than one
current and one voltage channel. For instance, if the user plans to use one voltage
channel and alternate measurements between two current channels, the relationship
between the voltage channel and each of the current channels can be characterized. The
amount of deskew required for each combination should be recorded for later use.
4.
Proceed directly to the power device measurement section by touching
Analysis
in the
menu bar, then
PowerMeasure
in the drop-down menu. Alternatively, touch one of the
Содержание 6000 SERIES
Страница 59: ...WR6K OM E Rev A ISSUED October 2003 57 4 Select a language 5 Click Next when the Wizard appears ...
Страница 60: ...Operator s Manual 58 ISSUED October 2003 WR6K OM E Rev A 6 Select Complete setup 7 Click Install ...
Страница 65: ...WR6K OM E Rev A ISSUED October 2003 63 Proper Orientation of Drive ...
Страница 131: ...WR6K OM E Rev A ISSUED October 2003 129 Example ...
Страница 141: ...WR6K OM E Rev A ISSUED October 2003 139 ...
Страница 236: ...Operator s Manual 234 ISSUED October 2003 WR6K OM E Rev A Example 6 ...
Страница 247: ...WR6K OM E Rev A ISSUED October 2003 245 In this figure the panel has been pasted onto the Excel sheet for comparison ...
Страница 261: ...WR6K OM E Rev A ISSUED October 2003 259 ...
Страница 286: ...Operator s Manual 284 ISSUED October 2003 WR6K OM E Rev A ...
Страница 304: ...Operator s Manual 302 ISSUED October 2003 WR6K OM E Rev A ...
Страница 307: ...WR6K OM E Rev A ISSUED October 2003 305 Convolve two signals ...
Страница 327: ...WR6K OM E Rev A ISSUED October 2003 325 How JitterTrack s Interval Error works when Clock Mode is selected ...
Страница 348: ...Operator s Manual 346 ISSUED October 2003 WR6K OM E Rev A ...
Страница 398: ...Operator s Manual 396 ISSUED October 2003 WR6K OM E Rev A Setting Up the Line Voltage and Current Signals ...
Страница 402: ...Operator s Manual 400 ISSUED October 2003 WR6K OM E Rev A ...