Operator's Manual
128
ISSUED: October 2003
WR6K-OM-E Rev A
on the software options loaded on your scope. See specifications..
3.
Touch inside the
Source1
field and select a source from the pop-up menu.
4.
Touch inside the
Operator1
field and select
Phistogram
from the
Select Math
Operator
menu.
5.
Touch the "Phistogram" tab, then touch inside the Slice Direction field and select
Horizontal
or
Vertical
slice from the pop-up menu.
6.
Touch inside the
Slice Center
field and enter a value, using the pop-up keypad.
7.
Touch inside the
Slice Width
field and enter a value, using the pop-up keypad.
Note:
You can use the front panel Adjust knobs to move the
Slice Center
line and the
Slice
Width
boundary lines.
Persistence Trace Range
This math operation has a field where you can enter the percent of the persistence trace
population to use in creating a new waveform.
Persistence Sigma
This math operation has a field where you can enter a scale, measured in standard deviations, by
which to create a new waveform.
HISTOGRAM PARAMETERS
fwhm
Full Width at Half Maximum
Definition:
Determines the width of the largest area peak, measured between bins on
either side of the highest bin in the peak that have a population of half the
highest's population. If several peaks have an area equal to the maximum
population, the leftmost peak is used in the computation.
Description:
First, the highest population peak is identified and the height of its highest bin
(population) determined (for a discussion on how peaks are determined see the
pks parameter Description:). Next, the populations of bins to the right and left
are found, until a bin on each side is found to have a population of less than
50% of that of the highest bin's. A line is calculated on each side, from the
center point of the first bin below the 50% population to that of the adjacent bin,
towards the highest bin. The intersection points of these lines with the 50%
height value is then determined. The length of a line connecting the intersection
points is the value for fwhm.
Содержание 6000 SERIES
Страница 59: ...WR6K OM E Rev A ISSUED October 2003 57 4 Select a language 5 Click Next when the Wizard appears ...
Страница 60: ...Operator s Manual 58 ISSUED October 2003 WR6K OM E Rev A 6 Select Complete setup 7 Click Install ...
Страница 65: ...WR6K OM E Rev A ISSUED October 2003 63 Proper Orientation of Drive ...
Страница 131: ...WR6K OM E Rev A ISSUED October 2003 129 Example ...
Страница 141: ...WR6K OM E Rev A ISSUED October 2003 139 ...
Страница 236: ...Operator s Manual 234 ISSUED October 2003 WR6K OM E Rev A Example 6 ...
Страница 247: ...WR6K OM E Rev A ISSUED October 2003 245 In this figure the panel has been pasted onto the Excel sheet for comparison ...
Страница 261: ...WR6K OM E Rev A ISSUED October 2003 259 ...
Страница 286: ...Operator s Manual 284 ISSUED October 2003 WR6K OM E Rev A ...
Страница 304: ...Operator s Manual 302 ISSUED October 2003 WR6K OM E Rev A ...
Страница 307: ...WR6K OM E Rev A ISSUED October 2003 305 Convolve two signals ...
Страница 327: ...WR6K OM E Rev A ISSUED October 2003 325 How JitterTrack s Interval Error works when Clock Mode is selected ...
Страница 348: ...Operator s Manual 346 ISSUED October 2003 WR6K OM E Rev A ...
Страница 398: ...Operator s Manual 396 ISSUED October 2003 WR6K OM E Rev A Setting Up the Line Voltage and Current Signals ...
Страница 402: ...Operator s Manual 400 ISSUED October 2003 WR6K OM E Rev A ...