Operator's Manual
254
ISSUED: October 2003
WR6K-OM-E Rev A
The next figure shows a part of the Excel workbook.
Here we see the gated waveform that has been created in Excel. The Mean parameter during the
region of interest (ROI) is placed in cell H3.
How Does this Work?
The amplitude of the signal is about 0.3 volts, and the screen height is 0.4 volts, as derived from
cells F7 and Fx. A threshold value for amplitude was calculated by placing 0.5 * (Fy – Fx) in cell
A4.
Remember that in the instrument the sources were defined to be A10 and B10. This means that
the first point on the waveform will be read into A10, and, since the waveform has 500 points, the
last point will be read into A510. The same holds true for F2 and column B, since F2 is assigned
as Source2, and data is defined to write into column B starting with cell B10.
To create the gating function in column C, the cell C10 was given the following formula:
IF ( ( B10 – B9) > $A$4, 1 – C9, C9). This was copied down the column. Column D, the output
column, is simply A * C.
The output was defined as cell H3.
The required mean in cell H3 is given by SUM (D10 : D509) / SUM (C10 : C509), for a 500 point
waveform.
Содержание 6000 SERIES
Страница 59: ...WR6K OM E Rev A ISSUED October 2003 57 4 Select a language 5 Click Next when the Wizard appears ...
Страница 60: ...Operator s Manual 58 ISSUED October 2003 WR6K OM E Rev A 6 Select Complete setup 7 Click Install ...
Страница 65: ...WR6K OM E Rev A ISSUED October 2003 63 Proper Orientation of Drive ...
Страница 131: ...WR6K OM E Rev A ISSUED October 2003 129 Example ...
Страница 141: ...WR6K OM E Rev A ISSUED October 2003 139 ...
Страница 236: ...Operator s Manual 234 ISSUED October 2003 WR6K OM E Rev A Example 6 ...
Страница 247: ...WR6K OM E Rev A ISSUED October 2003 245 In this figure the panel has been pasted onto the Excel sheet for comparison ...
Страница 261: ...WR6K OM E Rev A ISSUED October 2003 259 ...
Страница 286: ...Operator s Manual 284 ISSUED October 2003 WR6K OM E Rev A ...
Страница 304: ...Operator s Manual 302 ISSUED October 2003 WR6K OM E Rev A ...
Страница 307: ...WR6K OM E Rev A ISSUED October 2003 305 Convolve two signals ...
Страница 327: ...WR6K OM E Rev A ISSUED October 2003 325 How JitterTrack s Interval Error works when Clock Mode is selected ...
Страница 348: ...Operator s Manual 346 ISSUED October 2003 WR6K OM E Rev A ...
Страница 398: ...Operator s Manual 396 ISSUED October 2003 WR6K OM E Rev A Setting Up the Line Voltage and Current Signals ...
Страница 402: ...Operator s Manual 400 ISSUED October 2003 WR6K OM E Rev A ...