Operator's Manual
354
ISSUED: October 2003
WR6K-OM-E Rev A
In the setup of these filters, you have control of cutoff frequencies, transition region width, and
stop band attenuation.
FILTER SETUP
To Set Up a DFP Filter
1.
Touch
Math
in the menu bar then
Math Setup...
in the drop-down menu.
2.
Touch the
Fx
tab (F1
for example) for the math trace you want to
display your filtered waveform.
3.
Touch the
Single
function button
if you want to perform just one filtering function
on the trace, or touch the
Dual
function button
to perform math on, or apply
another filter to, the filter output.
4.
Touch inside the
Source1
field and select a source waveform from the pop-up menu.
5.
Touch inside the
Operaor1
field and select
Filter
from the pop-up menu. A mini-
dialog to set up the filter will open at right.
Note:
Other math choices in the
Operator1
menu include
Boxcar
, ERES, and interpolation. The
boxcar "filter" is a simple average taken over a user-specified number of points (the "length").
6.
Touch inside the
FIR/IIR
field and select finite or infinite response filter FIR (non-
recursive) filters require a limited number of multiplications, additions, and memory
locations. On the other hand, IIR (recursive) filters, which are dependent on previous
input or output values, in theory require an infinite number of each..
7.
Whether you selected FIR or IIR, touch inside the
Filter Kind
field and select a filtering
operation. Some choices are not available for IIR.
8.
If you selected FIR, touch inside the
Type
field and choose an FIR filter type. Then touch
inside the
Taps The number of coefficients.
data entry field and enter a value, using the
pop-up numeric keypad. Alternatively, you can touch the
Auto Length
checkbox; the
Taps
field disappears and the scope calculates the optimum number of coefficients. If
you selected IIR, touch inside the
Type
field and choose an IIR filter type. Then touch
inside the
Stages
data entry field and enter a value, using the pop-up numeric keypad.
Alternatively, you can touch the
Auto Length
checkbox; the
Stages
field disappears and
the scope calculates the optimum number of stages.
9.
Touch the
Frequencies
tab.
10.
Depending on the class (FIR/IIR) and kind of filter you selected, and whether or not Auto
Length is enabled, you can change the cutoff frequencies, transition width (edge width),
Содержание 6000 SERIES
Страница 59: ...WR6K OM E Rev A ISSUED October 2003 57 4 Select a language 5 Click Next when the Wizard appears ...
Страница 60: ...Operator s Manual 58 ISSUED October 2003 WR6K OM E Rev A 6 Select Complete setup 7 Click Install ...
Страница 65: ...WR6K OM E Rev A ISSUED October 2003 63 Proper Orientation of Drive ...
Страница 131: ...WR6K OM E Rev A ISSUED October 2003 129 Example ...
Страница 141: ...WR6K OM E Rev A ISSUED October 2003 139 ...
Страница 236: ...Operator s Manual 234 ISSUED October 2003 WR6K OM E Rev A Example 6 ...
Страница 247: ...WR6K OM E Rev A ISSUED October 2003 245 In this figure the panel has been pasted onto the Excel sheet for comparison ...
Страница 261: ...WR6K OM E Rev A ISSUED October 2003 259 ...
Страница 286: ...Operator s Manual 284 ISSUED October 2003 WR6K OM E Rev A ...
Страница 304: ...Operator s Manual 302 ISSUED October 2003 WR6K OM E Rev A ...
Страница 307: ...WR6K OM E Rev A ISSUED October 2003 305 Convolve two signals ...
Страница 327: ...WR6K OM E Rev A ISSUED October 2003 325 How JitterTrack s Interval Error works when Clock Mode is selected ...
Страница 348: ...Operator s Manual 346 ISSUED October 2003 WR6K OM E Rev A ...
Страница 398: ...Operator s Manual 396 ISSUED October 2003 WR6K OM E Rev A Setting Up the Line Voltage and Current Signals ...
Страница 402: ...Operator s Manual 400 ISSUED October 2003 WR6K OM E Rev A ...