166
ISSUED: October 2003
WR6K-OM-E Rev A
PARAMETER CALCULATIONS
Parameters and How They Work
Determining Top and Base Lines
Proper determination of the top and base reference lines is fundamental for ensuring correct
parameter calculations. The analysis begins by computing a histogram of the waveform data over
the time interval spanned by the left and right time cursors. For example, the histogram of a
waveform transitioning in two states will contain two peaks (see Figure 1). The analysis will
attempt to identify the two clusters that contain the largest data density. Then the most probable
state (centroids) associated with these two clusters will be computed to determine the top and
base reference levels: the top line corresponds to the top and the base line to the bottom
centroid.
Figure 1
Содержание 6000 SERIES
Страница 59: ...WR6K OM E Rev A ISSUED October 2003 57 4 Select a language 5 Click Next when the Wizard appears ...
Страница 60: ...Operator s Manual 58 ISSUED October 2003 WR6K OM E Rev A 6 Select Complete setup 7 Click Install ...
Страница 65: ...WR6K OM E Rev A ISSUED October 2003 63 Proper Orientation of Drive ...
Страница 131: ...WR6K OM E Rev A ISSUED October 2003 129 Example ...
Страница 141: ...WR6K OM E Rev A ISSUED October 2003 139 ...
Страница 236: ...Operator s Manual 234 ISSUED October 2003 WR6K OM E Rev A Example 6 ...
Страница 247: ...WR6K OM E Rev A ISSUED October 2003 245 In this figure the panel has been pasted onto the Excel sheet for comparison ...
Страница 261: ...WR6K OM E Rev A ISSUED October 2003 259 ...
Страница 286: ...Operator s Manual 284 ISSUED October 2003 WR6K OM E Rev A ...
Страница 304: ...Operator s Manual 302 ISSUED October 2003 WR6K OM E Rev A ...
Страница 307: ...WR6K OM E Rev A ISSUED October 2003 305 Convolve two signals ...
Страница 327: ...WR6K OM E Rev A ISSUED October 2003 325 How JitterTrack s Interval Error works when Clock Mode is selected ...
Страница 348: ...Operator s Manual 346 ISSUED October 2003 WR6K OM E Rev A ...
Страница 398: ...Operator s Manual 396 ISSUED October 2003 WR6K OM E Rev A Setting Up the Line Voltage and Current Signals ...
Страница 402: ...Operator s Manual 400 ISSUED October 2003 WR6K OM E Rev A ...