Operator's Manual
144
ISSUED: October 2003
WR6K-OM-E Rev A
The lowest-value bin with events is that with a sub-range of 3.0 to 3.1 V. As TTL high voltages
need to be greater than 2.5 V, the lowest bin is within the allowable tolerance. However, because
of its proximity to this tolerance and the degree of the bin's separation from all other values,
additional investigation may be required.
DSO Process
The instrument generates histograms of the parameter values of input waveforms. But first, you
must define the following:
•
The parameter to be histogrammed
•
The trace on which the histogram is to be displayed
•
The maximum number of parameter measurement values to be used in creating the
histogram
•
The measurement range of the histogram
•
The number of bins to be used
Some of these are pre-defined but can be changed. Once they are defined, the oscilloscope is
ready to make the histogram. The sequence for acquiring histogram data is as follows:
1.
Trigger
2.
Waveform acquisition
3.
Parameter calculations
4.
Histogram update
5.
Trigger re-arm
Содержание 6000 SERIES
Страница 59: ...WR6K OM E Rev A ISSUED October 2003 57 4 Select a language 5 Click Next when the Wizard appears ...
Страница 60: ...Operator s Manual 58 ISSUED October 2003 WR6K OM E Rev A 6 Select Complete setup 7 Click Install ...
Страница 65: ...WR6K OM E Rev A ISSUED October 2003 63 Proper Orientation of Drive ...
Страница 131: ...WR6K OM E Rev A ISSUED October 2003 129 Example ...
Страница 141: ...WR6K OM E Rev A ISSUED October 2003 139 ...
Страница 236: ...Operator s Manual 234 ISSUED October 2003 WR6K OM E Rev A Example 6 ...
Страница 247: ...WR6K OM E Rev A ISSUED October 2003 245 In this figure the panel has been pasted onto the Excel sheet for comparison ...
Страница 261: ...WR6K OM E Rev A ISSUED October 2003 259 ...
Страница 286: ...Operator s Manual 284 ISSUED October 2003 WR6K OM E Rev A ...
Страница 304: ...Operator s Manual 302 ISSUED October 2003 WR6K OM E Rev A ...
Страница 307: ...WR6K OM E Rev A ISSUED October 2003 305 Convolve two signals ...
Страница 327: ...WR6K OM E Rev A ISSUED October 2003 325 How JitterTrack s Interval Error works when Clock Mode is selected ...
Страница 348: ...Operator s Manual 346 ISSUED October 2003 WR6K OM E Rev A ...
Страница 398: ...Operator s Manual 396 ISSUED October 2003 WR6K OM E Rev A Setting Up the Line Voltage and Current Signals ...
Страница 402: ...Operator s Manual 400 ISSUED October 2003 WR6K OM E Rev A ...