- 70 -
RESERVED FOR THE INSTALLER
DATA DELETION AND PROGRAM SELECTION
WARNING
: operations must only be performed after contacting technical assistance.
After each operation the display shows
. By pressing the procedure is cancelled
and no changes are made.
Using the LCB Manager software, enter the instrument programming mode (see
PROGRAMMING
MODE
section in the LCB Manager manual) and proceed as follows:
CONSTANTS RESTORE
(does not erase the calibration): confirm
, use arrow keys to select
, set code 6935 and confirm.
PROGRAM SELECTION:
confirm
and use the arrow keys to select the desired program:
: basic program, setpoint management only.
: single product loading program.
Set the approval status
-
: not approved program;
-
: approved program, single interval (Dir. 2014/31/EU, art. 1)*;
-
: approved program, multi-interval (Dir. 2014/31/EU, art. 1)*;
-
: approved program, multiple range (Dir. 2014/31/EU, art. 1)*;
*
Contact technical assistance to request the proper manual and the correct procedures for
approval, indicating mandatory hardware code and serial number (see section
INSTRUMENT
COMMISSIONING
).
By confirming, the instrument is restored to default and data is erased.