- 11 -
THEORETICAL FULL SCALE
(default:
): the
system full scale
is given by one cell capacity multiplied by the
number of cells used. Example: 4 cells of 1000 kg
FULL SCALE = 1000 x 4 = 4000.
The instrument is supplied with a theoretical full scale value
corresponding to 10000.
To restore factory values, set 0 as full scale.
COMMUNICATION INTERFACE
WRITING
READING
Write the value in W1.
Send command 6000 to CMDR.
Send command 6001 to CMDR.
Read the value in R1.
SENSITIVITY
(default: 2.00000 mV/V): the
sensitivity
is a load cell rated parameter expressed in mV/V.
Set the average sensitivity value indicated on the load cells. It’s possible to set a value between
0.50000 and 7.00000 mV/V. Example: 4-cell system with sensitivity 2.00100, 2.00150, 2.00200,
2.00250; enter 2.00175, calculated as (2.00100 + 2.00150 + 2.00200 + 2.00250) / 4.
COMMUNICATION INTERFACE
WRITING
READING
Multiply the value by 100000.
Write the value in W1.
Send command 6008 to CMDR.
Send command 6007 to CMDR.
Read the value in R1.
Divide the value by 100000.
Example: to set
to 2.00175, write 200175 in W1 and send command 6008 to CMDR.