- 10 -
PROGRAMMING OF SYSTEM PARAMETERS
INSTRUMENT INFORMATION
The LCB Manager software shows, in the Information area of the main screen, some data
concerning the connected instrument (see the LCB Manager manual).
COMMUNICATION INTERFACE
Command sent to CMDR
Register to read
Register contents
1220
R1 (16 most significant bit)
software code
R1 (16 least significant bit)
firmware version
R2
hardware code
1221
R1 (16 most significant bit)
year of production
R1 (16 least significant bit)
serial number
R2
program code*
*0=BASE; 13=LOAD
THEORETICAL CALIBRATION
This function allows the load cell rated values to be set.
To perform the theoretical calibration, set in sequence the values of theoretical full scale, sensitivity
and division.
By modifying the theoretical full scale, the sensitivity or the division:
-
the real calibration is cancelled and the theoretical calibration only is considered valid;
-
the system parameters containing a weight value will be set to default values.