- 19 -
DISPLAY COEFFICIENT (BASE program)
By setting the coefficient
the display is changed accordingly.
If one of the inputs is set to
mode (see section
OUTPUTS AND INPUTS CONFIGURATION
(BASE program)
) when the input is closed the value will be displayed modified according to the
coefficient; when the input is opened the standard weight display will be restored.
: (max settable value: 99.9999; default: 1.0000) will have different meanings according to the
value set in
, i.e. the selected unit of measure. (see section
SETTING UNITS OF MEASURE
).
WARNING:
All other settings (setpoint, hysteresis, calibration ...) are expressed in weight
value. If you want to convert them to the new unit of measurement, perform one of the
following procedures for changing the system calibration.
The parameter
must remain set to 1.0000.
THEORETICAL CALIBRATION’S CHANGE FOR OTHER UNITS OF MEASURE
Set in the parameter
the F.SCALE value divided by the conversion coefficient from kg to
the new unit of measure.
Example: the 4 load cells of 1000 kg are placed under a scale for olive oil, which has a specific
gravity of 0.916 kg/l. Setting the F.SCALE = (4 x 1000) / 0.916 = 4367, the system works in liters of
olive oil.
REAL CALIBRATION’S CHANGE FOR OTHER UNITS OF MEASURE
Load a known quantity of product litres on the scale (equal to at least 50% of the maximum amount
that you must weigh) and enter in the parameter
, the product loaded value in litres.
COMMUNICATION INTERFACE
WRITING
READING
Write the value in W1.
Send command 6034 to CMDR.
Send command 6033 to CMDR.
Read the value in R1.
Write the enabling status* in W1.
Send command 6130 to CMDR.
Send command 6129 to CMDR.
Read the enabling status* in R1.
*1= enabled; 0= disabled