- 62 -
When the remote display is set to gross weight:
-
if the instrument displays the gross weight, the remote display shows the gross weight.
-
if the instrument shows the net weight, the remote display shows the net weight alternated
with the message
.
-
: transmission speed (2400, 4800, 9600, 19200, 38400, 115200; default: 9600).
-
: instrument address (from 1 to 99; default: 1).
-
: maximum transmission frequency (10 – 20 – 30 – 40 – 50 – 60 – 70 – 80 – 100 –
200 – 300; default: 10); to be set when the
transmission protocol is selected.
Maximum setting frequency
(
)
:
-
20 Hz with minimum baud rate 2400 baud.
-
40 Hz with minimum baud rate 4800 baud.
-
80 Hz with minimum baud rate 9600 baud.
-
100 Hz with minimum baud rate 19200 baud.
-
200 Hz with minimum baud rate 38400 baud.
-
300 Hz with minimum baud rate 38400 baud.
-
: delay in milliseconds which elapses before the instrument replies (from 0 to
200 ms; default: 0).
-
:
-
: no parity (default).
-
: even parity.
-
: odd parity.
-
: stop bit (1 – 2; default: 1).
FAST CONTINUOUS TRANSMISSION PROTOCOL
This protocol allows the continuous transmission of the weight at high update frequencies. Up to 300
strings per second are transmitted with a minimum transmission rate of 38400 baud.
Following communication modes availables (see section
SERIAL COMMUNICATION SETTINGS
):
: communication compatible with TX RS485 instruments.
: communication compatible with TD RS485 instruments.
If
is set, the following string is transmitted to PC/PLC:
xxxxxxCRLF
where:
xxxxxx
...... 6 characters of gross weight (48
57 ASCII)
CR
................ 1 character return to the start (13 ASCII)
LF
................ 1 character on new line (10 ASCII)