Integral XT
YAWE0028 / 24/06/2016
Maintenance
141
b)
High-temperature thermostats
Fault
Possible remedy
Master: Alarm message
Cool FloWX
(Equipment damage (lasting damage to the high
temperature valve)
)
1. Cooling water temperature above 80 °C and
longer than 8 seconds.
à
Restore correct
cooling water supply.
Contact LAUDA Service
Constant Temperature Equipment (
þ
2. Cooling water temperature above 85 °C.
à
Restore correct cooling water supply.
Contact LAUDA Service Constant Tempera-
ture Equipment.
3. At high-temperature valve temperature above
140 °C.
à
Restore correct cooling water sup-
ply.
Contact LAUDA Service Constant Tempera-
ture Equipment.
Master: Warning message
Hk uAluE too hok
(Equipment damage (lasting damage to the high
temperature valve)
)
1.
At high temperature valve temperature above
120 °C and more than 8 seconds.
à
Restore
correct cooling water supply.
Otherwise contact LAUDA Service Constant
Temperature Equipment (
þ
9.5).
Device enters the degassing mode (
þ
7.6.3).)
(Entry of cooling water in the hydraulic circuit by
defective condenser).
However, please note: If necessary, the device
performs a "permanently and automatic degassing"
by (
þ
7.6.3.2). This automatic process is not a
malfunction.
1. Contact LAUDA Service Constant Tempera-
ture Equipment.
Entry of heat transfer liquid in the cooling water
circuit by defective condenser.
1.
Note Suitable cooling water quality
(
þ
Otherwise there is a danger of corrosion!