Integral XT
YAWE0028 / 24/06/2016
Starting up
81
7.12.3 Interrupting, continuing or terminating the program (Hold, Continue, Stop)
Command
Programmer
Program 1
Status
−
After a program has been started by
pressing the
key, the command op-
tions
Hold
or
Stop
are shown.
−
Here, with the aid of the keys
or
and
the running program can be
paused with
Hold
or terminated with
Stop
.
−
Once the program has been terminated,
the device runs with the last setpoint set-
ting.
−
Continuation of a program paused with
Hold
occurs using
Continue
which is
obtained with
.
−
Also the standby key
stops the pro-
grammer. The pump, heater and cooling
unit are switched off.
Follow the safety information (
þ
−
After pressing the standby key
again,
the programmer returns to the previously
selected operating mode:
Pause or active operation depending on
what was previously selected.