Integral XT
154
Technical data
YAWE0028 / 11/11/2016
Table 3 Process thermostats
XT 750 (S)
XT 750 H(S)
XT 950 W(S)
XT 1590 W(S)
XT 1850 W(S)
Heater power /
Power consumption
400 V; 3/PE~50 Hz
kW
LWP 520:
5.3 / 7.8
LWP 522:
5.3 / 7.8
LWP 521:
5.3 / 7.8
---
LWP 532:
10.6 / 13.8
kW
LWP 552:
8.0 / 9.7
LWP 553:
8.0 / 9.7
LWP 554:
8.0 / 9.7
LWP 551:
8.0 / 13.8
LWP 533:
16.0 / 17.3
208-220 V; 3/PE~60 Hz kW
5.7 / 7.6
5.7 / 7.6
5.7 / 7.6
---
---
200 V; 3/PE~50/60 Hz kW
5.3 / 6.9
5.3 / 6.9
5.3 / 6.9
---
---
440-480 V; 3/PE~60 Hz
KW
---
---
---
7.0 / 16.6
14.0 / 20.8
400 V; 3/PE~50 Hz or KW
---
---
---
5.3 / 16.6 or
10.6 / 20.8 or
440-480 V; 3/PE~60 Hz KW
---
---
---
7.0 / 16.6
14.0 / 20.8
Surface loading (Heater)
400 V; 3/PE~50 Hz
W/cm²
LWP 520: 4.6
LWP 522: 4.6
LWP 521: 4.6
---
LWP 532: 4.6
W/cm²
LWP 552: 7.1
LWP 553: 7.1
LWP 554: 7.1
LWP 551: 7.1
LWP 533: 7.1
208-220 V; 3/PE~60 Hz W/cm²
5.1
5.1
5.1
---
---
200 V; 3/PE~50/60 Hz W/cm²
4.6
4.6
4.6
---
---
440-480 V; 3/PE~60 Hz W/cm²
---
---
---
6.1
6.1
400 V; 3/PE~50 Hz or
440-480 V; 3/PE~60 Hz
W/cm²
--- --- ---
4.6
or
6.1
4.6 or
6.1
Protection
IP21C
Pump type
Pressure pump
Pump
capacity
(water
20 °C)
Discharge pres-
sure maximum
bar
2.9 2.9 2.9 2.9 5.8
Flow rate max. L/min
45
45
45
45
90
Connections for consumers
Thread M30 x 1.5 (DN 20)
Thread M38 x 1.5
(DN 25)
Overall dimensions B x L x H
mm
460 x 550 x 1285
460 x 550 x 1285 460 x 550 x 1285
700 x 550 x 1600 700 x 550 x 1600
Weight
kg
155 160 160 280 250
Safety equipment
Class
III, FL suitable for flammable and non-flammable liquids
Protection class
Protection class I according to DIN EN 61140; VDE 0140-1
The refrigerating powers are reduced by about 320 watts when Pump Level 8 is selected instead of Pump Level 4. The
refrigerating powers are reduced by about 470 watts when Pump Level 8 is selected instead of Pump Level 2. With XT 1850
W(S) the refrigerating power is reduced by about 640 watts when Pump Level 8 is selected instead of Pump Level 4. With
XT 1850 W(S) the refrigerating power is reduced by about 940 watts when Pump Level 8 is selected instead of Pump Lev-
el 2.
Devices filled with Kryo 55 up to a maximum temperature of 200 °C. Devices filled with Ultra 350 up to a maximum tem-
perature of 300 °C.
Pump Level 2.
Water consumption for maximum refrigerating power
With pump connection link
Содержание Integral XT 150
Страница 2: ......
Страница 4: ......
Страница 164: ...Integral XT 162 YAWE0028 24 06 2016 Intentionally left blank ...
Страница 166: ......
Страница 167: ......