Integral XT
YAWE0028 / 24/06/2016
Starting up
99
7.15.4.2
Well-proven settings for control parameters and pump level for external control
Master controller
(external controller)
Slave control-
ler
(internal con-
troller)
Type of
device
Heat
transfer
liquid
Exter-
nal
applic-
ation
Kpe
EP
Tne
En
Tve
Eu
Tde
Ed
Prop_E
Eb
Xpf
iP
Pump
level
Press.
control
XT 150,
XT 250 W
KRYO 55
a)
4.0
300
246
24.0
20
5.0
3
--
KRYO 55
b)
1.5
300
246
24.0
20
7.0
1
1bar
KRYO 55
c)
0.7
100
84
8.0
20
7.0
4
--
XT 750 H,
XT 950 W
KRYO 55
d)
1.5
200
164
16.0
20
5.0
8
--
KRYO 55
e)
1.5
300
246
24.0
20
15.0
5
--
KRYO 55
f)
0.4
70
61
7.0
20
12.0
6
--
Description
a)
Glass double-shell reactor with 4 liters, uninsulated, thermostatic oil in the reactor, 2 x 1.5 m metal
corrugated hose D = 10 mm (internal).
b)
Glass double-shell reactor with 4 liters, uninsulated, thermostatic oil in the reactor, 2 x 6 m metal
corrugated hose D = 10 mm (internal), pressure control to P = 1 bar, bypass used.
c)
Load with low volume, low thermal capacity, low flow (cross-sections < 10 mm).
d)
Metal double-shell reactor, 17 liters of thermostatic oil in the reactor, 2 x 1.5 m metal corrugated hose
D = 20 mm
e)
Glass double-shell reactor with 4 liters, uninsulated, thermostatic oil in the reactor, 2 x 5 m metal
corrugated hose D = 10 mm (internal), bypass used.
f)
Load with low volume, low thermal capacity, low flow (cross-sections < 10 mm), bypass used.
7.15.5 Internal and external control parameter sets
If a thermostat is used for a number of applications, which always leads to a change of the control pa-
rameters, these control parameters (up to 9 sets) can be saved in the thermostat and activated again as
required.
Also saving is useful for finding the best control parameters; in this way external management of the
control parameters can be avoided.
There are 9 sets (each for internal and external sets of control parameters) saved at the factory.
In this menu the control parameters cannot be edited, they are only displayed.
•
With
Activate
the currently valid control parameters are used.
•
With
Upload actual
the actual ones are read in and saved (for later reuse).
•
With
Default
the set of control parameters saved at the works is loaded again (in this case the con-
trol parameters set by the customer are lost).