Integral XT
YAWE0028 / 24/06/2016
Preparations
35
Olives:
−
Push a hose onto the hose olive. Secure hoses against slippage with the aid of hose clips etc.
General notes:
−
Always ensure the largest possible passages in the external circuit. For a hose cross-section
that is too small
à
Temperature gradients occur between the device and external load due to
low volume flow.
−
Only operate the Integral XT when flow through the external load is possible.
−
Open any shut-off valves in the external loads.
−
Depending on the configuration of the load circuit, a venting valve can significantly simplify the
venting process. The venting valve should be positioned at the highest point of the hydraulic cir-
cuit (
þ
see the drawing on page 34).
−
Reactors for steam heating are not suitable as external loads, because they generally have an
area through which flow does not pass and in which vapor cushions can form.
−
If external control is to be used, provide a Pt100 probe in the external load.
−
Pay attention to pressure sensitive loads (e.g. glass apparatus) with a maximum
permissible operating pressure (
þ
−
Check whether the hoses for external loads have been mounted.
−
With outflow temperatures over 70 °C the supplied self-adhesive label (EZB
260) should be applied on the device at an easily visible point.
−
Do
not
carry out technical changes on the device!
+
−
The unit can be safely operated up to an ambient temperature of 40 °C.
−
An increased ambient temperature (above the reference temperature of 20 °C) re-
duces the cooling capacity and the minimum temperature that can be achieved.
−
With loads situated at a higher level and with the pump stopped and air seeping into
the thermostatic circuit (for example a not completely closed or defective venting
valve), then even with enclosed circuits, the external volume can run empty.
à
Danger that the process thermostat will overflow!
−
Install a dirt trap if the complete heat transfer system on the customer side is not
guaranteed to be dirt free.
Connection of the cooling water
Note that the following conditions apply for the connection of the cooling water supply:
Cooling water pressure (feed - outlet)
maximum 10 bar overpressure
Differential pressure (feed - outlet)
minimum 3.0 bar
Cooling water temperature
10 to 15 °C recommended,
10 to 30 °C admissible (with power restrictions)
Cooling water quantity
see Technical Data (
þ
Cooling water hose for connection to the
device
minimum 13 mm (up to XT 950 W)
minimum 19 mm (XT 1590 W(S), XT 1850 W(S))