Integral XT
62
Starting up
YAWE0028 / 24/06/2016
7.9.4 Pressure control
Alternatively to the 8 pump power levels, a mode with pressure control is provided which facilitates a
very effective supply of pressure-sensitive glass reactors with a maximum permissible pressure rating.
Command
Pressure control
−
Open the device parameter menu using
the soft key
Menu .
−
Change
from
Pump
à
to
Pressure control
.
−
The settings window opens.
−
Enter the required pressure. The setting
option with the power levels (
þ
7.9.3) is
thus switched off.
−
In the overview display P
set
and P
int
are
displayed.
−
If a too high pressure is set which the
pump cannot achieve, then the pump
operates at its power limit (Level 8).
7.9.5 Maximum pressure control
With the operation of double-shell vessels or other pressure-sensitive applications, the maximum sys-
tem pressure must be set (reduced).
+
−
This setting does not replace the function of the component-tested safety valve
(
þ
Bursting of the external consumer
Scalding, frostbite
•
For consumers with a maximum permissible opera-
ting pressure below the maximum pressure of the
pump, use a safety valve for protection. This safety
valve must be installed in the outlet of the device
Help
Menu Screen
T
set
T
fix
6
7
8
9
0
1
2
3
4
5
Input set pressure (0 = off)
0,
00
Min: 0,0 Max: 7,0 bar