Integral XT
YAWE0028 / 24/06/2016
Starting up
59
−
Select desired position with the cursor
keys
(black background).
−
With the soft key
Edit open the
window shown on the left.
−
Enter fixed temperature setpoint as de-
scribed above and accept into the list
with
or cancel with
.
−
Select and accept values from the list of
fixed settings as described above for the
"Recent setpoints".
7.9.2 Displaying the actual external temperature
With all Integral XT Thermostats an external temperature probe can be connected, which for exam-
ple.....
1. ...can be used as an independent temperature measurement channel,
2. ...can be used as the controlled variable for the bath temperature in applications with a no-
ticeable temperature gradient (between the internal bath temperature and an external load).
The setup is described in Section (
þ
7.9.6). With the function described in the following, you
only change over the display.
+
−
External actual temperatures can also be read in by interface modules (
þ
Connection of the external Pt100 to the Lemo socket 10S.
Bench-top device
Floor-standing device
Contact on
socket 10S
1
2
3
4
+
+
-
-
I
U
U
I
Current circuit
Voltage circuit
Voltage circuit
Current circuit
Pt100
DIN EN 60751
−
Plug: 4-pole Lemosa for Pt100 connection (Order No. EQS 022).
−
Use screened connecting leads. Connect screen to plug case.
Help
Menu Screen
T
set
T
fix
6
7
8
9
0
1
2
3
4
5
Enter new setpoint:
123.
45
Min: -40.00°C Max:202.00°C