Integral XT
YAWE0028/ 11/11/2016
Safety information
7
Explanation of signs:
Danger:
This sign is used where there may be injury to person-
nel if a recommendation is not followed accurately or is
disregarded.
Note:
Here special attention is drawn to some aspect. May
include reference to danger.
Reference
Refers to other information in different sections.
1
Safety information
According to Paragraph 14 of the operational safety decree (BetrSichV)
1
, the device is a system which
requires supervision. (Classification according to the Pressure Equipment Directive 97/ 23 / EC: Catego-
ry I). Before being put into operation, the system must be subjected to inspection for siting, for ascer-
taining that the device is in order and for correct functioning. A certificate must be issued regarding this
inspection, documenting the extent and the result of the inspection.
1
The national regulations of the respective country in which the system is sited must be followed.
1.1
General safety information
A process thermostat is used to heat, cool and circulate heat transfer liquids as specified. Hazards arise
from this due to high or low temperatures, excess pressures, fire and the general hazards due to the
application of electrical energy.
The user is largely protected by the application of the relevant standards.
Further hazard sources can arise from the type of material for which the temperature is to be stabilized,
e.g. by the exceeding or undercutting certain temperature thresholds or by the fracture of the container
and reaction with the heat transfer liquid.
It is not possible to include all possibilities. They remain essentially subject to the judgment and respon-
sibility of the operator.
The devices may only be used as intended, that is as described in this operating manual. This includes
operation by instructed specialist personnel.
The devices are not designed for use in medical applications in accordance with DIN EN 60601-1 or
IEC 601-1.
Use restriction
On the EMC standard DIN EN 61326-1:
Devices in Emissions Class A can only be operated on power grids that are not con-
nected to residential areas!
Classification in accordance with EMC requirements
Device
Immunity
Emissions class
Customer power supply
Integral XT high-
temperature ther-
mostat
Type 2 in accordance
with
DIN EN 61326-1
Emissions Class B
in accordance with
CISPR 11
Worldwide
No limitation