Integral XT
YAWE0028 / 24/06/2016
Starting up
73
−
The temperature measurement device
shows the true temperature value (with
glass thermometers take the correction
into account where applicable).
−
Change the display in the adjacent win-
dow to the true value with cursor or soft
keys and accept with
or End ,
−
or quit the window with
without mak-
ing changes.
7.10.7 Restoring the works setting of the internal temperature-probe offset
If the offset has been misadjusted unintentionally, the works setting can be restored with this function.
Command
Default
−
Open the device parameter menu via
the soft key Menu .
−
With the cursor keys continue to:
à
Settings
à
Calibration
à
intern Pt100
à
Default
.
−
The window shown adjacent appears.
−
Confirm selection with
.
−
Confirm the control dialog on the right
with
or cancel with
.
−
Return to the measurement window
with End or
.
Help
Menu Screen
T
set
T
fix
6
7
8
9
0
1
2
3
4
5
Temperature value of the ref.
temp. measurement device:
20,
15
Pump Menu
End
T
set
T
fix
internal Pt100
external Pt100
Calibration
Default
Pump Menu
End
T
set
T
fix
Confirm input!
Enter key: Continue
Escape key: Cancel
Содержание Integral XT 150