7.10.5 Setting the volume of the acoustic signals
The LAUDA Integral XT Thermostats signal alarms as a dual-tone acoustic signal and warnings as a
continuous tone.
Command
Sounds
−
Open the device parameter menu via the
soft key Menu .
−
With the cursor keys continue to:
à
Settings
à
Basic Settings
à
Sounds Master
or
Sounds Command
.
−
Select either
Alarm
or
Warning
.
−
Example on left:
Alarm
is set to loud.
−
Select the desired volume with
or
.
−
Accept selection with
or End or
quit the window with
without making
changes.
7.10.6 Entering the offset of the internal temperature probe
If, during checking with a calibrated reference thermometer probe, e.g. from the LAUDA DigiCal Series,
a deviation is found, then the offset (i.e. the additive part of the characteristic) of the internal measuring
chain can be adjusted with the following function. The reference thermometer must be fitted according
to the details on the calibration certificate in the outflow.
Command
Calibration
−
Open the device parameter menu via
the soft key Menu .
−
With the cursor keys continue to:
à
Settings
à
Calibration
à
Internal Pt100
à
Calibration
.
−
The window shown on the left appears.
−
Confirm selection with
.
Pump Menu
End
T
set
T
fix
internal Pt100
external Pt100
Calibration
Default
Pump Menu
End
T
set
T
fix
loud
medium
low
off
Alarm