Lake Shore Model 330 Autotuning Temperature Controller User’s Manual
Options and Accessories
6-1
CHAPTER 6
OPTIONS AND ACCESSORIES
6.0 GENERAL
This chapter covers Model 330 Temperature Controller model numbers (Paragraph 6.1), Options (Paragraph
6.2), and Accessories (Paragraph 6.3).
6.1 MODELS
Two Model 330 independent sensor inputs can be variously configured with certain combinations of the
following input types. Users can reconfigure any of the input types except for the optional thermocouple input.
Model
Channel A Sensor
Channel B Sensor
330-11
Silicon Diode
Silicon Diode
330-12
Silicon Diode
Platinum Resistor
330-13
Silicon Diode
GaAlAs Diode
330-22
Platinum Resistor
Platinum Resistor
330-23
Platinum Resistor
GaAlAs Diode
330-33
GaAlAs Diode
GaAlAs Diode
330-41 Thermocouple
Silicon
Diode
330-42 Thermocouple
Platinum
Resistor
330-43 Thermocouple
GaAlAs
Diode
330-44 Thermocouple
Thermocouple
330-51 Thermocouple
Platinum
Resistor
330-53 Thermocouple
GaAlAs
Diode
330-55 Thermocouple
Thermocouple
Use the model numbers above to specify input type (e.g., 330-12 for a Model 330 with one Silicon Diode input
and one Platinum input). The standard 330 heater output is 25 watts. Add a suffix of W50 to the order number
for a unit factory-configured for 50 watts operation, or the user can do so by moving a jumper within the unit.
6.2 OPTIONS
MODEL
NUMBER
DESCRIPTION OF MODEL 330 OPTIONS
8000
Precision Option, Floppy Disk.
Consists of breakpoint pairs from a Sensor Precision
Calibration loaded on a floppy disk in ASCII format for Customer downloading.
8001
Precision Option, Factory Installed.
Provides custom factory programming of a specific
sensor calibration curve. The Precision Option improves combined sensor/instrument
accuracy to within ±0.1 K or better over the calibrated temperature range of the sensor. Data
is stored in a memory chip (NOVRAM). Requires the use of a calibrated sensor.
8002-05
Precision Option, Field Installation.
For field installation of the precision option for Model 330
owners. When ordering, specify your instrument serial number and calibrated sensor model
and serial number. A new NOVRAM will be sent for Customer installation.