Lake Shore Model 330 Autotuning Temperature Controller User’s Manual
Remote Operation
4-19
4.3.6 Curve
Commands
Curve Commands allow users to verify existing factory-added curves or enter or delete user-defined curves
over the interface.
ACOMP
Set Room Temperature Compensation for Channel A.
Input:
ACOMP 0
or
ACOMP 1
Returned:
Nothing
Remarks:
Effective only with the Model 330-4X Thermocouple Version. Select temperature
compensation parameter:
0
= off,
1
= on.
ACOMP?
Room Temperature Compensation for Channel A Query.
Input:
ACOMP?
Returned:
0
or
1
Remarks:
Effective only with the Model 330-4X Thermocouple Version. Returns current room
temperature compensation status:
0
= off,
1
= on.
ACUR
Assign Curve Number for Channel A.
Input:
ACUR XX
Returned:
Nothing
Remarks:
Enter integer from 0 through 31 for Channel A. Table 3-1 lists sensor curve numbers.
ACUR?
Curve Number for Channel A Query.
Input:
ACUR?
Returned:
XX
(an integer from 00 to 31)
Remarks:
Returns the currently selected sensor curve number for Channel A. Table 3-1 lists sensor
curve numbers.
ATYPE?
Channel A Input Type Query.
Input:
ATYPE?
Returned:
SI
,
PT
,
AS
,
TC
, or
ER
Remarks:
Returns input type for Channel A:
SI
= silicon diode,
PT
= platinum,
AS
= GaAlAs,
TC
= thermocouple,
ER
= error (improper switch setting).
BCOMP
Set Room Temperature Compensation for Channel B.
Input:
BCOMP 0
or
BCOMP 1
Returned:
Nothing
Remarks:
Effective only with the Model 330-4X Thermocouple Version. Select temperature
compensation parameter:
0
= off,
1
= on.
BCOMP?
Room Temperature Compensation for Channel B Query.
Input:
BCOMP?
Returned:
0
or
1
Remarks:
Effective only with the Model 330-4X Thermocouple Version. Returns current room
temperature compensation status:
0
= off,
1
= on.
BCUR
Assign Curve Number for Channel B.
Input:
BCUR XX
Returned:
Nothing
Remarks:
Enter an integer from 0 through 31 for Channel B. Table 3-1 lists sensor curve numbers.