Lake Shore Model 330 Autotuning Temperature Controller User’s Manual
3-14
Operation
Zone 01
0 K
Zone 02
Zone 03
Zone 04
Zone 05
Zone 06
Zone 07
Zone 08
Zone 09
Zone 10
Setpoint:
Setpoint:
Setpoint:
Setpoint:
Setpoint:
Setpoint:
Setpoint:
Setpoint:
Setpoint:
Setpoint:
Heater
Low
0 K
Heater
Low
Heater
High
Heater
High
Heater
High
Heater
High
Heater
High
Heater
High
Heater
High
Heater
High
P –
100
I
– 20
D –
00
10 K
20 K
30 K
40 K
50 K
60 K
70 K
80 K
90 K
100 K
Default Zone Settings
P
– 90
I
– 20
D
– 00
P
– 80
I
– 20
D
– 00
P
– 70
I
– 20
D
– 00
P
– 60
I
– 20
D
– 00
P
– 50
I
– 20
D
– 00
P
– 40
I
– 20
D
– 00
P
– 30
I
– 20
D
– 00
P
– 20
I
– 20
D
– 00
P
– 10
I
– 20
D
– 00
3.4.5
Power Up (PUP) Configuration
Store a Model 330 Power Up (PUP) configuration to ensure that
it powers up to a user-defined state after power down.
Parameters including heater range, setpoint, gain, reset, units,
and curve number store in non-volatile memory and are
retained even when the line cord is disconnected.
To view PUP status, hold the
Enter
key for about 5 seconds to
see the display to the right.
“On” updates power up settings when the user changes them
via the front panel or over the remote interface. “On” is the
default PUP. If left on, the instrument powers up in the same
configuration it was powered down in.
“Off” disables updates to the power up memory. The instrument powers up in the configuration it was in when
the power up feature was turned off.
To change PUP status, hold the
Enter
key and press the
t
key to toggle the PUP status between On and Off.
Sample
OFF
Control
LOC
K
A
K
B