Lake Shore Model 330 Autotuning Temperature Controller User’s Manual
5-6
Service & Calibration
5.8 ERROR
MESSAGES
On power up, the Model 330 checks internal memory. There are two potential error messages.
The first error,
Er01
, indicates an unsuccessful attempt to write and then read the internal non-volatile RAM. This error is not
user-correctable. Consult the factory.
The second error,
Er02
, indicates an unsuccessful attempt to read the internal non-volatile RAM for the Model
ID. Sometimes, initializing memory corrects this error. To initialize memory, press and hold both
Escape
and
Units
for about 20 seconds. Release when the power up sequence begins. Perform this operation only under
extreme circumstances; it erases any user-defined curves in memory.
There are two sensor input error messages. If an input signal from the sensor exceeding full scale is applied
to the input leads,
OL
indicates the overload.
If no signal or a wrong polarity signal exists at the input leads,
Er27
indicates a Zero Error for Channel A or
Er28
for Channel B. The Model 330 displays dashes “
- - -
” if the
input switches are improperly set.
Finally,
Er30
appears if the measured heater output does not match the predicted output. The controller turns
the heater off. Check heater resistance, test for shorts in the heater wiring, then turn the heater on again. If the
error message returns, consult the factory.
Table 5-1. Sensor Input Setup
1
2
3
PT
AS
TC
Sensor
Sensor
DIP Switch
DIP Switch
Settings
Settings
Display When
Display When
Input Type
Input Type
Key is Pressed
Key is Pressed
Sensor
Sensor
Units
Units
Input
Input
Range
Range
Standard
Standard
Curve(s)
Curve(s)
Silicon Diode
(DT-400
Series)
Volts
0 to 2.5
2 Curve10
4 Curve 10 SoftCal
Thermocouple
(compensation
on)
Millivolts
15 to 15
1
2
3
PT
AS
TC
Platinum
RTD
Ohms
0 to 300
3 DIN Curve 43760
1
2
3
PT
AS
TC
GaAlAs
Diode
Volts
0 to 6.0
None
1
2
3
PT
AS
TC
10 to 10
6 AuFe 0.07% vs. Chromel
7 AuFe 0.03% vs. Chromel
8 Type E (chromel-constantan)
9 Type K (chromel-alumel)
10 Type T (copper-constantan)
1
2
3
PT
AS
TC
Thermocouple
(compensation
off)
Millivolts
*
*
DIP Switch S1 is for Channel A and S2 is for Channel B.
If the internal DIP switches are improperly set, the
display will read Er when the
Input Type
Input Type
key is
pressed. The normal front panel display will show
dashes - - - - to indicate improper DIP switch setting.
Improper Switch Setting