82
2.7 Electrical Unit
REGIUS MODEL 190 Service Manual Ver.1.00 2004.11.01
6.
Remove a
fi
xing screw, and pull out the MCB2 from board’s
support legs (5pcs).
7.
Carry out reassembly in the reverse order to that described
above.
•
Remarks for Installation
•
Set the rotary SW (SW4) and DIP SW (SW3) on the new
board according to the note taken in step 3.
•
Check that the DIP SW(SW2)’s all dips are set to OFF
(set to the left) before turning on the power.
•
Update the
fi
rmware using the REGIUS Console.
MCB2
Fixing Screw
SW2
Board’s Support
Legs
Содержание REGIUS 190
Страница 7: ...VI REGIUS MODEL 190 Service Manual Ver 1 00 2004 11 01...
Страница 13: ...Blank...
Страница 27: ...xx REGIUS MODEL 190 Service Manual Ver 1 00 2004 11 01...
Страница 28: ...Chap 1 Overview...
Страница 29: ...Blank Page...
Страница 43: ...16 1 3 Layout of Controlling Components REGIUS MODEL 190 Service Manual Ver 1 00 2004 11 01...
Страница 45: ...Blank Page...
Страница 52: ...Chap 2 Disassembly and Reassembly...
Страница 53: ...Blank Page...
Страница 118: ...Chap 3 Maintenance Adjustment...
Страница 119: ......
Страница 153: ...126 3 9 Adjusting the Position of Insertion Guide Switch SW4 REGIUS MODEL 190 Service Manual Ver 1 00 2004 11 01...
Страница 154: ...Chap 4 Troubleshooting...
Страница 155: ......
Страница 194: ...A Appendix...
Страница 195: ......
Страница 218: ...191 A 5 Parts Layout Diagram REGIUS MODEL 190 Service Manual Ver 1 00 2004 11 A 5 2 SCB2 System Control Board...
Страница 219: ...192 A 5 Parts Layout Diagram REGIUS MODEL 190 Service Manual Ver 1 00 2004 11 A 5 3 ADB2 A D Board...
Страница 220: ...0676 55030 B 0412TE...