4.4 Error Code
REGIUS MODEL 190 Service Manual Ver.1.00 2004.11.01
4.4.7
SCB2 Error Codes
Code
Medium Classifi-
cation
Detailed Clas-
sification
Countermeasures, etc.
10000
Memory error
SCB2 replacement or CF card
replacement
11000
Hardware error
SCB2 replacement
12000
Internal communi-
cation error
Mechanical
control com-
munication
High voltage ON
Timeout
Con
fi
rm that there was no continuous
insertion.
• Open and close the door and then
restore. If restoration is not possi-
ble, restart.
12001
Internal error
Mechanical
control com-
munication
Laser ON Timeout
Open and close the door and then
restore. If restoration is not possible,
restart. Following restart, con
fi
rm the
following.
LD peripherals (LD power supply
and optical unit)
Check the wiring on the board.
12002
Internal error
Mechanical
control com-
munication
Mech Ready Time-
out
Open and close the door and then
restore. If restoration is not possible,
restart.
12003
Internal error
Mechanical
control com-
munication
Standby Timeout
Open and close the door and then
restore. If restoration is not possible,
restart.
12010
Internal error
Mechanical
control com-
munication
Resent counts
Over
fl
ow
Open and close the door and then
restore. If restoration is not possible,
restart.
12011
Internal error
Mechanical
control com-
munication
CRC Erorr counts
Over-
fl
ow
Open and close the door and then
restore. If restoration is not possible,
restart.
12200
Internal error
Network com-
munication
Net Initialize Time-
out
Open and close the door and then
restore. If restoration is not possible,
restart. If the problem then reoccurs,
con
fi
rm whether the JM is operating.
Also con
fi
rm that the device is con-
nected to the net.
12201
Internal error
Network com-
munication
Net Ready Timeout
Open and close the door and then
restore. If restoration is not possible,
restart.
12202
Internal error
Network com-
munication
Put Shading Time-
out
Open and close the door and then
restore. If restoration is not possible,
restart.
12203
Internal error
Network com-
munication
Status Check Time-
out
Open and close the door and then
restore. If restoration is not possible,
restart.
12204
Internal error
Network com-
munication
Barcode Seek Tim-
eout
Open and close the door and then
restore. If restoration is not possible,
restart.
Содержание REGIUS 190
Страница 7: ...VI REGIUS MODEL 190 Service Manual Ver 1 00 2004 11 01...
Страница 13: ...Blank...
Страница 27: ...xx REGIUS MODEL 190 Service Manual Ver 1 00 2004 11 01...
Страница 28: ...Chap 1 Overview...
Страница 29: ...Blank Page...
Страница 43: ...16 1 3 Layout of Controlling Components REGIUS MODEL 190 Service Manual Ver 1 00 2004 11 01...
Страница 45: ...Blank Page...
Страница 52: ...Chap 2 Disassembly and Reassembly...
Страница 53: ...Blank Page...
Страница 118: ...Chap 3 Maintenance Adjustment...
Страница 119: ......
Страница 153: ...126 3 9 Adjusting the Position of Insertion Guide Switch SW4 REGIUS MODEL 190 Service Manual Ver 1 00 2004 11 01...
Страница 154: ...Chap 4 Troubleshooting...
Страница 155: ......
Страница 194: ...A Appendix...
Страница 195: ......
Страница 218: ...191 A 5 Parts Layout Diagram REGIUS MODEL 190 Service Manual Ver 1 00 2004 11 A 5 2 SCB2 System Control Board...
Страница 219: ...192 A 5 Parts Layout Diagram REGIUS MODEL 190 Service Manual Ver 1 00 2004 11 A 5 3 ADB2 A D Board...
Страница 220: ...0676 55030 B 0412TE...