184
A.2 DIP Switch Setting
REGIUS MODEL 190 Service Manual Ver.1.00 2004.11.01
A.2 DIP Switch Setting
A.2.1 DIP Switches on the MCB2 (Mechanical Control Board)
•
Operation Mode Setting
•
SIO CH0 Setting
•
SIO CH1 Setting
•
Cassette-Fall Detection
•
Durability Test Mode (expanded mode)
SW2-1
SW2-2
Setting Mode
Description
OFF
OFF
Normal operation
mode
Normal operation mode that allows the REGIUS 190 to perform
the normal reading sequence.
ON
OFF
Stand-alone test
mode
Connecting the REGIUS 190 and the PC with serial cable,
enables to implement each individual operation command.
OFF
ON
Continuous opera-
tion mode
Enables to repeat the reading operation of the REGIUS 190 with-
out cassette.
ON
ON
Durability test mode
A mode to perform durability test of the cassette. Difference from
the normal mode is; subscanning is made only for stroke of
5cm(up & down).
SW2-3
SW2-4
Setting Mode
Description
OFF
OFF
Bar code reader
Enables the bar code reader on the REGIUS 190.
ON
OFF
No input/output
Disables the I/O operation of the speci
fi
ed port. Reading opera-
tion is performed by default cassette size.
OFF
ON
Debug I/O-1
Output the debug information from this port. Do not read the bar
code, and operation is performed by default cassette size.
ON
ON
Debug I/O-2
Output the debug information from CN12 port. Do not read the
bar code, and operation is performed by default cassette size.
SW2-5
SW2-6
Setting Mode
Description
OFF
OFF
SCB2 communica-
tion
Enables the communication with SCB2.
ON
OFF
No input/output
Disables the I/O operation of the speci
fi
ed port.
OFF
ON
Debug I/O
Output the debug information from this port. For it does not com-
municate with SCB2, registration mode is
fi
xed to bar code regis-
tration, erase speed is
fi
xed to the maximum.
ON
ON
SW2-7
Setting Mode
Description
OFF
Enabled
Enables the detection of cassette fall.
ON
Disabled
Disables the detection of cassette fall.
SW2-8
Setting Mode
Description
OFF
Expanded
Eject the cassette at every cycle.
Содержание REGIUS 190
Страница 7: ...VI REGIUS MODEL 190 Service Manual Ver 1 00 2004 11 01...
Страница 13: ...Blank...
Страница 27: ...xx REGIUS MODEL 190 Service Manual Ver 1 00 2004 11 01...
Страница 28: ...Chap 1 Overview...
Страница 29: ...Blank Page...
Страница 43: ...16 1 3 Layout of Controlling Components REGIUS MODEL 190 Service Manual Ver 1 00 2004 11 01...
Страница 45: ...Blank Page...
Страница 52: ...Chap 2 Disassembly and Reassembly...
Страница 53: ...Blank Page...
Страница 118: ...Chap 3 Maintenance Adjustment...
Страница 119: ......
Страница 153: ...126 3 9 Adjusting the Position of Insertion Guide Switch SW4 REGIUS MODEL 190 Service Manual Ver 1 00 2004 11 01...
Страница 154: ...Chap 4 Troubleshooting...
Страница 155: ......
Страница 194: ...A Appendix...
Страница 195: ......
Страница 218: ...191 A 5 Parts Layout Diagram REGIUS MODEL 190 Service Manual Ver 1 00 2004 11 A 5 2 SCB2 System Control Board...
Страница 219: ...192 A 5 Parts Layout Diagram REGIUS MODEL 190 Service Manual Ver 1 00 2004 11 A 5 3 ADB2 A D Board...
Страница 220: ...0676 55030 B 0412TE...