4
1.1 Principal Specifications
REGIUS MODEL 190 Service Manual Ver.1.00 2004.11.01
Exposure cycle time
The time between
insertion and removal of
the cassette.
Approximately 40 seconds maximum
(for an image size of 14" x 14" and at 175
ı
)
Processing Capacity
90 or more images per hour (for an image size of 14" x 14" and at 175
ı
)
Sampling Pitch
3 types, i.e.,Super High Precision Mode (43.75µm), High Precision mode
(87.5
ı
) and Normal mode (175
ı
)
Max. Reading Size
(mm)
352.0 x 428.5 (14" x 17" cassette)
Max. Number of Pixels
7080 x 9480
QR Value
3 types, i.e., Low Sensitivity (QR value = 125), Standard Sensitivity (QR value =
250), and High Sensitivity (QR value = 500)
• A QR value of 200 is de
fi
ned for 1,535 steps at an exposure of 2.58 x 10
-7
C/kg
(=1mR) and 12-bit output for the output signal value.
Dynamic Range (read)
Normal D range (4.0 digits: PMT output current range = 100 nA to 1 mA)
Digital Density Steps
12 bit (4,096 steps)
Содержание REGIUS 190
Страница 7: ...VI REGIUS MODEL 190 Service Manual Ver 1 00 2004 11 01...
Страница 13: ...Blank...
Страница 27: ...xx REGIUS MODEL 190 Service Manual Ver 1 00 2004 11 01...
Страница 28: ...Chap 1 Overview...
Страница 29: ...Blank Page...
Страница 43: ...16 1 3 Layout of Controlling Components REGIUS MODEL 190 Service Manual Ver 1 00 2004 11 01...
Страница 45: ...Blank Page...
Страница 52: ...Chap 2 Disassembly and Reassembly...
Страница 53: ...Blank Page...
Страница 118: ...Chap 3 Maintenance Adjustment...
Страница 119: ......
Страница 153: ...126 3 9 Adjusting the Position of Insertion Guide Switch SW4 REGIUS MODEL 190 Service Manual Ver 1 00 2004 11 01...
Страница 154: ...Chap 4 Troubleshooting...
Страница 155: ......
Страница 194: ...A Appendix...
Страница 195: ......
Страница 218: ...191 A 5 Parts Layout Diagram REGIUS MODEL 190 Service Manual Ver 1 00 2004 11 A 5 2 SCB2 System Control Board...
Страница 219: ...192 A 5 Parts Layout Diagram REGIUS MODEL 190 Service Manual Ver 1 00 2004 11 A 5 3 ADB2 A D Board...
Страница 220: ...0676 55030 B 0412TE...