106
3.3 Verification of Initializing Sequence
REGIUS MODEL 190 Service Manual Ver.1.00 2004.11.01
5.
Move the Elevator Unit (Cassette Receive Unit) to the HP.
1)
Click "Move to HP" of "Cassette receive Unit".
2)
Check that "Move to Cassette receive Unit...OK" is
shown on "Information : ".
6.
Check the initialization sequence of Cassette receive Unit.
1)
Click "Initialize" of "Cassette receive Unit".
2)
Check that "Cassette receive Unit Initialize...OK" is
shown on "Information : ".
7.
Move the Cassette receive Unit to the transport standby
position.
1)
Click "Transport pos." of "Cassette receive Unit".
2)
Check that "Cassette receive Unit at Transport-
Standby...OK" is shown on "Information : ".
8.
Check the initialization sequence of sub scanning unit.
1)
Click "Initialize" of "Sub scanning".
2)
Check that "Sub scanning Unit Initialize...OK" is shown
on "Information : ".
Содержание REGIUS 190
Страница 7: ...VI REGIUS MODEL 190 Service Manual Ver 1 00 2004 11 01...
Страница 13: ...Blank...
Страница 27: ...xx REGIUS MODEL 190 Service Manual Ver 1 00 2004 11 01...
Страница 28: ...Chap 1 Overview...
Страница 29: ...Blank Page...
Страница 43: ...16 1 3 Layout of Controlling Components REGIUS MODEL 190 Service Manual Ver 1 00 2004 11 01...
Страница 45: ...Blank Page...
Страница 52: ...Chap 2 Disassembly and Reassembly...
Страница 53: ...Blank Page...
Страница 118: ...Chap 3 Maintenance Adjustment...
Страница 119: ......
Страница 153: ...126 3 9 Adjusting the Position of Insertion Guide Switch SW4 REGIUS MODEL 190 Service Manual Ver 1 00 2004 11 01...
Страница 154: ...Chap 4 Troubleshooting...
Страница 155: ......
Страница 194: ...A Appendix...
Страница 195: ......
Страница 218: ...191 A 5 Parts Layout Diagram REGIUS MODEL 190 Service Manual Ver 1 00 2004 11 A 5 2 SCB2 System Control Board...
Страница 219: ...192 A 5 Parts Layout Diagram REGIUS MODEL 190 Service Manual Ver 1 00 2004 11 A 5 3 ADB2 A D Board...
Страница 220: ...0676 55030 B 0412TE...