186
A.2 DIP Switch Setting
REGIUS MODEL 190 Service Manual Ver.1.00 2004.11.01
A.2.2 DIP Switches on the ADB2 (Analogue/Digital Board)
•
Input Mode Selection
•
Log AMP Input Selection
•
Setting of Reference Current
•
Setting of Polygon CLK
•
High Voltage Setting Mode Selection
•
High Voltage ON/OFF Selection
•
Offset Adjustment Mode Selection
SW1-1
Setting Mode
Description
OFF
Manual
Various settings are made by manual. (use SW1-2 ~ 6)
ON
Remote
Various settings can be controlled by SCB2.
SW1-2
Setting Mode
Description
OFF
Ref. Curr.
Uses reference current for the input to Log AMP.
ON
PMT
Uses PMT current for the input to Log AMP.
SW1-3
SW1-4
Setting Mode
Description
OFF
OFF
-50uA
Set the reference current at -50uA.
ON
OFF
-500nA
Set the reference current at -500nA.
OFF
ON
-50nA
Set the reference current at -50nA.
ON
ON
0A
Set the reference current at 0A.
SW1-5
SW1-6
Setting Mode
Description
OFF
OFF
Setting-1
Set the Polygon CLK to 659.3472Hz. (setting except for mam-
mography)
ON
OFF
Setting-2
Set the Polygon CLK to 329.6236Hz. (setting except for mam-
mography)
OFF
ON
Setting-3
Set the Polygon CLK to 454.7224Hz. (setting for mammography)
ON
ON
Setting-4
Set the Polygon CLK to 227.3612Hz. (setting for mammography)
SW2
Setting Mode
Description
A
Auto.
High voltage setting is controlled by the data sent from SCB2.
M
Manual
High voltage setting is controlled by VR2.
SW3
Setting Mode
Description
OFF
OFF
High voltage OFF.
ON
ON
High voltage ON.
SW4
Setting Mode
Description
A
Auto.
Setting of offset adjustment is controlled by the data sent from SCB2.
M
Manual
Setting of offset adjustment is controlled by VR1.
Содержание REGIUS 190
Страница 7: ...VI REGIUS MODEL 190 Service Manual Ver 1 00 2004 11 01...
Страница 13: ...Blank...
Страница 27: ...xx REGIUS MODEL 190 Service Manual Ver 1 00 2004 11 01...
Страница 28: ...Chap 1 Overview...
Страница 29: ...Blank Page...
Страница 43: ...16 1 3 Layout of Controlling Components REGIUS MODEL 190 Service Manual Ver 1 00 2004 11 01...
Страница 45: ...Blank Page...
Страница 52: ...Chap 2 Disassembly and Reassembly...
Страница 53: ...Blank Page...
Страница 118: ...Chap 3 Maintenance Adjustment...
Страница 119: ......
Страница 153: ...126 3 9 Adjusting the Position of Insertion Guide Switch SW4 REGIUS MODEL 190 Service Manual Ver 1 00 2004 11 01...
Страница 154: ...Chap 4 Troubleshooting...
Страница 155: ......
Страница 194: ...A Appendix...
Страница 195: ......
Страница 218: ...191 A 5 Parts Layout Diagram REGIUS MODEL 190 Service Manual Ver 1 00 2004 11 A 5 2 SCB2 System Control Board...
Страница 219: ...192 A 5 Parts Layout Diagram REGIUS MODEL 190 Service Manual Ver 1 00 2004 11 A 5 3 ADB2 A D Board...
Страница 220: ...0676 55030 B 0412TE...