196
④
After setting, the screen displays
as right. Click [Set].
[Set]
⑤
Click [OK] to finish adjustment
of index error.
[OK]
Note: 1. Repeat the checking steps to measure the Index Error (i angle). If the Index
Error cannot meet the requirement; user should check whether the three steps of the
adjustment and the collimation are right. Then set again according to the requirement.
2. If Index Error still not meets the requirement after the repeated operation, the
instrument should be returned to factory for inspection and repair.
● The vertical angle displayed in zero point setting has not been compensated or
modified, only for reference during setting.
15.7 HORIZONTAL AXIS ERROR COMPENSATION ADJUSTMENT
As the horizontal axis error only affects the angle of sight, it can be only
confirmed through observing the target the height of which is obviously lower or
higher than the instrument.
To avoid the influence of sight axis, user must have an associated adjustment
before adjusting sight axis.
It is unnecessary to collimate the prism or the target plane to ascertain the
transverse axis error. Therefore user is enabled to launch this adjustment at any time.
Select a recognizable point which is rather far away from the instrument, and much
higher or lower than the instrument. Make sure it can be precisely collimated twice.