4-26 Operation Method
RC04-PCR-LA
• Status Signal
The status signal outputs while the period of time set for the
T2, T3 and T4 (if T2=T4=0, this case applies T3) in the
Power Line Abnormality Simulation or the setting time of
address in the Sequence Operation.
When a status signal is speci
fi
ed to be ON, the signal “L” is
output to the SEQ STAT OUT terminal (BNC connector) on
the rear of the PCR-LA series. When a status signal is speci-
fi
ed to be OFF, the signal “H” is output. “H” is approximately
5 V, and the “L” is approximately 0 V.
The BNC connectors are isolated from the INPUT terminal
board and the OUTPUT terminal board of the PCR-LA series.
Note that the common line of each signal for the PEAK INIT
IN, SEQ TRG OUT, and SEQ STAT OUT are not isolated,
because they are internally shared in the unit of PCR-LA
series. And the internal circuits of Slot’s are also not isolated.
There is a slight time difference (approx. 100
μ
s) between the
status-signal outputs and the actual output changes.
Fig.4-11
SEQ STAT OUT Terminal and Its Circuit
• The status signal may output when any change occurred in
the parameter setting under the Power Line Abnormality Sim-
ulation or in the sequence setting under the Sequence Opera-
tion.
DESCRIPTION
SLOT 1
SLOT 2 (GPIB)
SLOT 3
SLOT 4
SEQ STAT
OUT terminal
Upper rear part of PCR-LA
AC Power Supply
Internal circuit of PCR-LA
NOTE