4-8 Operation Method
RC04-PCR-LA
• Note that when voltage data is read from the memory in the
AC + DC mode for output, output voltage changes as follows:
An AC voltage set value (Vrms) and DC voltage set value
(Vdc) are independently stored in the memory. When they are
read, the AC voltage value (Vrms) is output slightly faster
(approximately 30
μ
s) than the DC voltage value (Vdc).
Therefore, output voltage changes as follows:
Example
ADRS 1:
Vrms = 0V,
Vdc = +100V
ADRS 2:
Vrms = 100V,
Vdc = 0V
ADRS 3:
Vrms = 0V,
Vdc = +100V
Fig.4-2
Changes in the Output Voltage in Reading
Data from Memory
In the actual case, the duration of 30
μ
s shown in the above
fi
gure is very short in comparison with each ADRS time. Out-
put voltage changes like pulse.
NOTE
30
μ
s
30
μ
s
ADRS3
ADRS2
ADRS1
(100 V)-
(241 V)-
0 V-
Vrms (ADRS1)
+ Vdc (ADRS1)
Vrms (ADRS3)
+ Vdc (ADRS3)
Vrms (ADRS2)
+ Vdc (ADRS2)
Vrms (ADRS3) + Vdc (ADRS2)
Vrms (ADRS2) + Vdc (ADRS1)