RC04-PCR-LA
Operation Method 4-37
Crest factor = Peak value / rms value
For a sine wave, the crest factor is 1.41.
For a commercial power line voltage waveform, the sine wave
crest factor should be 1.41. However, since the peak of the
actual waveform is suppressed, the crest factor is between 1.2
and 1.4.
• A sine wave, peak-clipped waveform, and user-de
fi
ned wave-
form can be arbitrarily assigned to waveform banks W01 to
W14.
However, a user-de
fi
ned waveform is only temporarily cap-
tured in the PCR-LA power supply, and is erased when the
POWER switch of the power supply is turned OFF. (Note that
the sine waves and peak-clipped waveforms are backed up in
the power supply.)
Special Waveform Setting Procedure
A special waveform can be set only when the OUTPUT key is OFF in
the Home Position.
• Special waveform setting is available only in the AC, AC-S, or AC
+ DC mode.
• There are 15 waveform banks, W00 to W14, but W00 is for read
only.
1. Press the ESC key to call the Home Position.
2. Press the W-EDIT (SHIFT, WAVE) key.
The PCR-LA power supply will enter the key-lock mode automat-
ically.
In the Initial Setup Status, the following display (
fi
rst display)
appears.
3. Press the MENU key to select the display to change the
waveform banks to be displayed.
Three displays are provided. These are selected in turn whenever
the MENU key is pressed.
NOTE
(Sine waves have
already been set to
waveform banks
W00 through W04.)
S I N
S I N
S I N
S I N
W 0 1
W 0 2
W 0 3
W 0 4
F1
F2
F3
F4
F5
Содержание RC04-PCR-LA
Страница 1: ...RC04 PCR LA Part No Z1 002 902 IA002963 Jan 2007 OPERATION MANUAL REMOTE CONTROLLER PCR LA Series...
Страница 10: ...1 4 General RC04 PCR LA...
Страница 78: ...7 4 Specifications RC04 PCR LA 7 2 Dimensions Fig 7 1 Dimension of RC04 PCR LA 40 MAX 55 128 180 Unit mm...