RC04-PCR-LA
Operation Method 4-17
7. Use the +/- (SHIFT, 0) key to set the voltage change start
polarity (POL).
Voltage change start polarity (POL) is positive (+) in the Initial
Setup Status. To change the polarity to negative (-), press the +/-
(SHIFT, 0) key.
The set polarity will be displayed in the “POL” area of the
Remote Controller. For the voltage regulation start polarity
(POL), see the Description on page 4-18.
8. Press the MENU key to call the second display.
Set each item in the same way as in steps 5 to 7.
9. Press the MENU key to display the third display.
Set each item in the same way as in steps 5 to 7.
• The display is switched in the order of the
fi
rst to third and the
execution mode display whenever the MENU key is pressed.
• To return to the previous display, press the (SHIFT, MENU) keys.
10.Press the MENU key to display the execution mode, which
is the 4th display.
11.Set the OUTPUT key to ON.
If the T2 and T4 set values are any values other than 0 ms, this
step may be performed prior to step 4.
12.Press the F1 (RUN) key. This starts the simulation of power
line abnormality.
During power line abnormality simulation, the “RUN” indication
on the Remote Controller lights up and “1” in the S-MODE area
on the PCR-LA power supply control panel lights up.
S i m u l a t i o n
S t o p
R U N
S T O P
F1
F2
F3
F4
F5
(Second display)
0 m s
0 m s
0 . 0 V
T 4
T 5
V ( T 3 )
F1
F2
F3
F4
F5
(Third display)
∞
P R T
F1
F2
F3
F4
F5